R&S ESU
Instrument Functions
Analyzer Mode
4.156
Operating Manual 1302.6163.12 - 03
The reference value for the phase noise measurement can be defined with
REF
POINT LEVEL
,
REF POINT FREQUENCY
and
REF POINT LVL OFFSET
to differ
from that of the reference marker.
Remote command:
--
PH NOISE
ON | OFF
The
PH NOISE ON/OFF
softkey switches on/off the phase noise measurement.
Switching on is performed by means of the
PHASE NOISE
softkey and is only nec-
essary when the phase noise measurement has been switched off in the submenu.
Remote command:
CALC:DELT1:FUNC:PNO ON
CALC:DELT1:FUNC:PNO:RES?
REF POINT
LEVEL
The
REF POINT LEVEL
softkey activates an entry box for the input of a reference
level other than the reference marker level. The function is identical to that of the
softkey with the same name in the
MARKER
menu (
MKR
).
Remote command:
CALC:DELT1:FUNC:FIX:RPO:Y -10dB
REF POINT LVL
OFFSET
The
REF POINT LVL OFFSET
softkey activates an entry box for the input of an
additional level offset for the phase noise calculation.
This level offset is set to 0 dB on when the
REFERENCE FIXED
or
PHASE NOISE
function is enabled.
Remote command:
CALC:DELT:FUNC:FIX:RPO:Y:OFFS 10dB
REF POINT
FREQUENCY
The
REF POINT FREQUENCY
softkey activates an entry box for the manual input
of a reference frequency for the
REFERENCE FIXED
or
PHASE NOISE
function.
Remote command:
CALC:DELT1:FUNC:FIX:RPO:X 10.7MHz
PEAK SEARCH
The PEAK SEARCH sets the reference point level for delta marker 2 in the selected
measurement window to the peak of the selected trace.
Remote command:
CALC:DELT:FUNC:FIX:RPO:MAX
AUTO PEAK
SEARCH
The AUTO PEAK SEARCH softkey activates an automatic peak search for the refer-
ence fixed marker 1 at the end of each particular sweep.
This function may be used for tracking of a drifting source during a phase noise
measurement. The delta marker 2, which shows the phase noise measurement
result, keeps the delta frequency value. Therefore the phase noise measurement
leads to reliable results in a certain offset although the source is drifting. Only if the
marker 2 is reaching the border of the span, the delta marker value is adjusted to be
within the span. In these cases, choose a larger span.
Remote command:
CALC:DELT:FUNC:PNO:AUTO ON | OFF
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...