Operating Manual 1302.6163.12 - 03
6.25
R&S ESU
Remote Control – Description of Commands
CALCulate Subsystem
Characteristics:
*RST value: -
SCPI: conform
Mode:
A
CALCulate<1|2>:LIMit<1...8>:CLEar[:IMMediate]
This command deletes the result of the current limit check for all limit lines.
Up to 8 limit lines can be defined at the same time. The numeric suffixes <1|2> of CALCulate indicate
the measurement window.
This command is an event and therefore has no *RST value.
Example:
"CALC:LIM:CLE"
'Deletes the result of the limit check in screen A.
Characteristics:
*RST value: -
SCPI: conform
Mode:
A
CALCulate<1|2>:LIMit<1...8>:COMMent
<string>
This command defines a comment for the limit line selected (max. 40 characters).
Up to 8 limit lines can be defined at the same time. The comment is independent from the
measurement window.
Example:
"CALC:LIM5:COMM 'Upper limit for spectrum'"
'Defines the comment for limit line 5.
Characteristics:
*RST value: blank comment
SCPI: device-specific
Mode:
R, A
CALCulate<1|2>:LIMit<1...8>:COPY
1 to 8 | <name>
This command copies one limit line onto another one.
Up to 8 limit lines can be defined at the same time. The command is independent of the measurement
window.
Parameter:
1 to n ::= number of the new limit line or
<name> ::= name of the new limit line given as a string
Example:
"CALC:LIM1:COPY 2"
'Copies limit line 1 to line 2.
"CALC:LIM1:COPY 'FM2'"
'Copies limit line 1 to a new line named 'FM2'.
Characteristics:
*RST value: --
SCPI: device-specific
Mode:
R, A
This command is an event and therefore has no *RST value and no query.
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...