R&S ESU
Instrument Functions
Basic Settings
4.258
Operating Manual 1302.6163.12 - 03
ANNOTATION
ON/OFF
The
ANNOTATION ON/OFF
softkey switches the displaying of frequency informa-
tion on the screen on and off.
• ON: Frequency information is displayed.
• OFF: Frequency information is not outputted to the display. This can be used for
example to protect confidential data.
Remote command:
DISP:ANN:FREQ ON
DATA ENTRY
OPAQUE
The
DATA ENTRY OPAQUE
softkey sets the data entry windows to opaque. This
means that entry windows are underlaid with the background color for tables.
Remote command:
--
DEFAULT
COLORS 1/
DEFAULT COL
ORS 2
The
DEFAULT COLORS 1/2
softkeys restores the default settings for brightness,
color tint and color saturation for all display screen elements.
The color schemes have been selected to give optimum visibility of all picture ele-
ments at an angle of vision from above or below.
DEFAULT COLORS 1
is active in
the default setting of the instrument.
Remote command:
DISP:CMAP:DEF1
DISP:CMAP:DEF2
DISPLAY PWR
SAVE
The
DISPLAY PWR SAVE
softkey is used to switch on/off the power-save mode for
the display and to enter the time for the power-save function to respond. After the
elapse of this time the display is completely switched off, i.e. including backlighting.
The power-save mode is configured as follows:
• The first keystroke activates the power-save mode and opens the editor for the
response time. The response time is entered in minutes between 1 and 6 minutes
and is confirmed by
ENTER
.
• The power-save mode is deactivated by pressing the key again.
On leaving the menu with the power-save mode in the activated state, the softkey is
highlighted in color on returning to the menu and opens again the editor for the
response time. Pressing again the key switches off the power-save mode.
Remote command:
DISP:PSAV ON
DISP:PSAV:HOLD 15
This mode is recommended for saving the TFT display especially when the instru-
ment is exclusively operated in remote control.
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...