Instrument Functions
Analyzer Mode
4.210
Operating Manual 1302.6163.12 - 03
CHANNEL
BANDWIDTH
The
CHANNEL BANDWIDTH
softkey opens a window for selecting the measure-
ment channel bandwidth.
The default setting is 14 kHz.
The specified channel bandwidth allows the optimal setting of the measurement
parameters of the R&S ESU using
ADJUST SETTINGS
.
Remote command:
SENS:POW:ACH:BWID 14kHz
ADJUST
SETTINGS
The
ADJUST SETTINGS
softkey adapts the span to the channel bandwidth
selected.
For the carrier/noise ratio measurement, the span is set to:
4 × channel ban measurement margin
The adjustment is performed once; if necessary, the setting can be changed later
on.
Remote command:
SENS:POW:ACH:PRES CN | CN0
4.8.13.6
Measurement of the AM Modulation Depth
MODULATION
DEPTH
The
MODULATION DEPTH
softkey switches on the measurement of the AM modu-
lation depth. An AM-modulated carrier is required on the screen for ensuring correct
operation.
The level value of
MARKER 1
is taken as the carrier level. When this function is acti-
vated,
MARKER 2
and
MARKER 3
are automatically set symmetrically to the carrier
on the adjacent peak values of the trace as delta markers and
MARKER 2
is acti-
vated for the entry.
When the position of MARKER 2 (delta) is changed, MARKER 3 (delta) is moved
symmetrically with respect to the reference marker (MARKER 1).
If the data entry is activated for MARKER 3 (
MARKER 1 2 3 4
softkey), the latter can
be moved for fine adjustment irrespective of MARKER 2.
The R&S ESU calculates the power at the marker positions from the measured lev-
els. The AM modulation depth is calculated from the ratio between the power values
at the reference marker and at the delta markers. When the powers of the two AM
side bands are unequal, the mean value of the two power values is used for AM
modulation depth calculation.
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...