Operating Manual 1302.6163.12 - 03
6.149
R&S ESU
Remote Control – Description of Commands
MMEMory Subsystem
Parameter:
<file_name> := DOS file name
Example:
"MMEM:STOR:SPUR 'D:\TEST.ASC'"
Characteristics:
*RST value: -
SCPI: device-specific
Mode:
A
This command is an event and therefore has no *RST value and no query.
MMEMory:STORe<1|2>:STATe
1,<file_name>
This command stores the current device settings in a specified file name.
The file name includes indication of the path and may also include the drive name. The numeric suffix
in STORe<1|2> is irrelevant with this command.
For a detailed discription of the selectable items refer to the command
Parameter:
<file_name> = DOS file name without extension
Example:
"MMEM:STOR:STAT 1,'TEST'"
Characteristics:
*RST value: -
SCPI: conform
Mode:
R, A
This command is an event and therefore has no *RST value and no query.
MMEMory:STORe<1|2>:TRACe
1 to 3,<file_name>
This command stores the selected trace in a file with ASCII format. The file format is described in
chapter
“ASCII FILE EXPORT” on page 4.135
The numeric suffix at STORe<1|2> indicates the measurement window (screen A or B).
The decimal separator (decimal point or comma) for floating-point numerals contained in the file is
defined with the
FORMat:DEXPort:DSEParator
command.
Parameter:
1 to 3 := selected measurement curve (trace)
<file_name> := file name in compliance to DOS conventions, including indication
of the path and the drive name
Example:
"MMEM:STOR2:TRAC 3,'F:\TEST.ASC'"
'Stores trace 3 from screen B in the file TEST.ASC on a memory stick.
Characteristics:
*RST value: -
SCPI: device-specific
Mode:
R, A
This command is an event and therefore has no *RST value and no query.
For Spurious Measurement an extended file format is used:
Different language versions of evaluation programs may require different handling
of the decimal point. Using the
DECIM SEP
softkey, you can thus choose between
the delimiters '.' (decimal point) and ',' (comma).
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...