CALCulate Subsystem
6.38
Operating Manual 1302.6163.12 - 03
CALCulate<1|2>:LIMit<1...8>:LOWer:SPACing
LINear | LOGarithmic
This command selects linear or logarithmic interpolation for the lower limit line.
Example:
"CALC:LIM:LOW:SPAC LIN"
Characteristics:
*RST value: LIN
SCPI: device-specific
Mode:
A
CALCulate<1|2>:LIMit<1...8>:LOWer:THReshold
<numeric_value>
This command defines an absolute threshold value for limit lines with relative y-axis scaling
independently of the measurement window. The absolute threshold value is used in the limit check as
soon as it exceeds the relative limit value.
The unit must correspond to the unit selected with
CALC:LIM:UNIT
(except dB which is not allowed).
If no unit is indicated, the unit defined with
CALC:LIM:UNIT
is automatically used (exception: dBm
instead of dB).
The units DEG, RAD, S, HZ, PCT are not available.
Example:
"CALC:LIM2:LOW:THR -35DBM"
'Defines an absolute threshold value for limit line 2.
Characteristics:
*RST value: -200 dBm
SCPI: device-specific
Mode:
A
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...