Operating Manual 1302.6163.12 - 03
4.159
R&S ESU
Instrument Functions
Analyzer Mode
NEW SEARCH
The
NEW SEARCH
softkey starts a new peak search and enters the results in the
peak list.
Remote command:
INIT;*WAI;
CALC:MARK:FUNC:FPE 10;
CALC:MARK:FUNC:FPE:COUN?;
CALC:MARK:FUNC:FPE:Y?;
CALC:MARK:FUNC:FPE:X?
SORT MODE
FREQ | LEVEL
The
SORT MODE FREQ/LEVEL
softkey defines the position of the peak values in
the list:
• FREQ: sorting in ascending order of frequency values (time values if span = 0)
• LEVEL: sorting according to level
Remote command:
CALC:MARK:FUNC:FPE:SORT X;
PEAK
EXCURSION
With level measurements, the
PEAK EXCURSION
softkey allows the minimum
amount to be entered by which a signal must decrease or increase in order to be
recognized as a maximum by the peak search function.
Values between 0 dB and 80 dB may be entered, the resolution being 0.1 dB
Remote command:
CALC:MARK:PEXC 6dB
LEFT LIMIT |
RIGHT LIMIT
The
LEFT LIMIT
and
RIGHT LIMIT
softkeys define the vertical lines F1/F2 in the fre-
quency domain (span > 0) and T1/T2 in the time domain (span = 0) between which
the search is carried out.
If only one line is active, the F1/T1 line is used as the lower limit; the upper limit is
the stop frequency. If F2/T2 is also active, it defines the upper limit.
Remote command:
CALC:MARK:X:SLIM:LEFT 1MHZ
CALC:MARK:X:SLIM:RIGH 10MHZ
CALC:MARK:X:SLIM ON
THRESHOLD
The
THRESHOLD
softkey defines a horizontal threshold line which represents the
lower limit of the peak search level range.
Remote command:
CALC:THR -20dBm
CALC:THR ON
PEAK LIST OFF
The
PEAK LIST OFF
softkey switches off the table with the search results.
Remote command:
--
PEAK LIST
EXPORT
The
PEAK LIST EXPORT
softkey stores the content of the marker peak list in ASCII
format to the specified file. The format of the decimal point is defined by the
DECIM
SEP
softkey.
Remote command:
MMEM:STOR:PEAK 'C:\filename.txt'
DECIM SEP
The
DECIM SEP
softkey selects the decimal separator between '.' (decimal point)
and ',' (comma) for the function
PEAK LIST EXPORT
.
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...