6.44
Operating Manual 1302.6163.12 - 03
CALCulate<1|2>:MARKer<1...4>:X:SLIMits:RIGHt
0 to MAX (frequency | sweep time)
This command sets the right limit of the search range for markers and delta markers in the selected
measurement window. Depending on the x-axis domain the indicated value defines a frequency (span
> 0) or time (span = 0). The function is independent of the selection of a marker, i.e. the numeric suffix
in MARKer<1...4> is irrelevant.
If the time domain power measurement is active, this command limits the evaluation range to the trace.
Example:
"CALC:MARK:X:SLIM ON"
'Switches the search limit function on for screen A.
"CALC:MARK:X:SLIM:RIGH 20MHz"
'Sets the right limit of the search range in screen A to 20 MHz.
Characteristics:
*RST value: - (is set to the right diagram border on switching on search limits)
SCPI: device-specific
Mode:
R, A
CALCulate<1|2>:MARKer<1...4>:COUNt
ON | OFF
This command switches on or off the frequency counter at the marker position in the selected
measurement window. The count result is queried with
CALCulate:MARKer:COUNt:FREQuency?
.
Frequency counting is possible only for one marker at a time for each measurement window. If it is
activated for another marker, it is automatically de-activated for the previous marker.
It should be noted that a complete sweep must be performed after switching on the frequency counter
to ensure that the frequency to be measured is actually reached. The synchronization to the sweep
end required for this is possible only in single-sweep mode.
Example:
"INIT:CONT OFF"
'Switches to single-sweep mode.
"CALC:MARK ON"
'Switches on marker 1 in screen A.
"CALC:MARK:COUN ON"
'Switches on the frequency counter for marker 1.
"INIT;*WAI"
'Starts a sweep and waits for the end.
"CALC:MARK:COUN:FREQ?"
'Outputs the measured value in screen A.
Characteristics:
*RST value: OFF
SCPI: device-specific
Mode:
A
The function is only available if the search limit for marker and delta marker is
switched on (
CALC:MARK:X:SLIM ON
).
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...