Operating Manual 1302.6163.12 - 03
4.263
R&S ESU
Instrument Functions
Basic Settings
• The
TRANSDUCER
softkey opens a submenu for entering the correction
characteristics for transducers. For details refer to section
• The
GENERAL SETUP
softkey opens a submenu for all the general settings such
as GPIB address, date and time as well as the configuration of the device
interfaces.
FIRMWARE OPTIONS
can be installed under this menu item. For
details refer to section
• The
SYSTEM INFO
softkey opens a submenu for displaying the hardware
configuration of the instrument, the switching cycle statistics and system
messages. For details refer to section
• The
SERVICE
softkey opens a submenu in which special device functions and
system information can be selected for servicing. The password required for
service functions can be entered in this submenu. For details refer to section
.
• The
FIRMWARE UPDATE
opens a submenu to update the firmware version, to
restore the firmware and to update the firmware path. For details refer to section
“Firmware Update” on page 4.296
• The
NOISE SRC ON/OFF
softkey switches on and off the voltage supply for an
external noise source. For details refer to section
• The
IF SHIFT
softkey opens a submenu to activate or deactivate 1
st
IF shifting.
For details refer to
• The
RECEIVER PRESET
softkey sets the receiver mode as preset mode. For
“PRESET RECEIVER” on page 4.297
• The
ANALYZER PRESET
softkey sets the analyzer mode as preset mode. For
“PRESET ANALYZER” on page 4.297
• The
SERVICE FUNCTIONS
softkey enables additional special settings for
servicing and troubleshooting. It is available after entering the corresponding
password under the
SERVICE
softkey.
4.9.3.1
External Reference
The R&S ESU can use the internal reference source or an external reference
source as frequency standard from which all internal oscillators are derived. A 10
MHz crystal oscillator is used as internal reference source. In the default setting
(internal reference), this frequency is available as output signal at rear-panel con-
nector REF OUT, e.g. to synchronize other instruments to the reference of the
R&S ESU.
In the setting
REFERENCE EXT
, the connector REF IN is used as input connector
for an external frequency standard. In this case all internal oscillators of the
R&S ESU are synchronized to the external reference frequency.
REFERENCE
FREQUENCY
As of firmware version 4.4x, instead of the
REFERENCE INT / EXT
softkey, the firm-
ware provides the
REFERENCE FREQUENCY
softkey. This softkey opens a sub-
menu to configure the reference more precisely.
To maintain the setup of a test system, the reference settings do not change if a pre-
set is performed.
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...