R&S ESU
Instrument Functions
Basic Settings
4.278
Operating Manual 1302.6163.12 - 03
Factors
- Selection of factors for the subrange
The permissible transducer factors for the marked subrange can be selected in a
selection box. Only factors matching with the unit of the set and fully covering the
selected subrange are permissible.
After each change of range limits, R&S ESU thus checks the factor list and, if
required, rebuilds it.
After reducing the start frequency or increasing the stop frequency of a range it may
happen that the factors defined for this range no longer fully cover the range. These
factors are deleted for this range when the transducer factor table is opened next
time.
A maximum of 4 transducer factors can be switched on at the same time in each
subrange. If none of them is switched on, 0 dB is assumed as a factor for the whole
subrange.
Remote command:
--
SAVE TRD SET
The
SAVE TRD SET
softkey saves the changed table in a file on the internal hard
disk. If a transducer name with the same name already exists, a corresponding
query is performed beforehand:
If the saved set is switched on, the new values will be used immediately.
Remote command:
--
(executed automatically)
4.9.3.5
Programming the Interface Configuration and Time Setup
The
GENERAL SETUP
softkey opens a submenu in which the general instrument
parameters can be set up. In addition to the configuration of the digital interfaces
(
IECBUS, COM
), the date and time may be entered.
The current settings are displayed in tabular form on the display screen where they
may be edited.
GENERAL
SETUP
!
!
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...