Operating Manual 1302.6163.12 - 03
4.283
R&S ESU
Instrument Functions
Basic Settings
Serial Interface Configuration
COM
INTERFACE
The
COM INTERFACE
softkey activates the
COM INTERFACE
table for entry of the
serial interface parameters.
The following parameters can be configured in the table:
Baud
– Data transmission rate
The R&S ESU supports baud rates between 110 and 128000 baud. The default set-
ting is 9600 baud.
Remote command:
SYST:COMM:SER:BAUD 9600
Bits –
Number of data bits per word
For the transmission of text without special characters, 7 bits are adequate. For
binary data as well as for text with special characters, 8 bits must be selected
(default setting).
Remote command:
SYST:COMM:SER:BITS 7
Baud rate
data
transmission
rate
Bits
number of data bits
Parity
bit parity check
Stop bits
number of stop bits
HW-Handshake
hardware handshake protocol
SW-Handshake
software handshake protocol
Owner
assignment to the measuring instrument or computer
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...