Operating Manual 1302.6163.12 - 03
6.39
R&S ESU
Remote Control – Description of Commands
CALCulate Subsystem
6.5.3.4
CALCulate:LIMit:UPPer Subsystem
The CALCulate:LIMit:UPPer subsystem defines the upper limit line.
CALCulate<1|2>:LIMit<1...8>:UPPer[:DATA]
<numeric_value>,<numeric_value>...
This command defines the values for the upper limit lines.
Up to 8 limit lines can be defined at the same time. This command is independent of the measurement
window.
The number of values for the CONTrol axis and for the corresponding UPPer and/or LOWer limit line
have to be identical. Otherwise default values are entered for missing values or not necessary values
are deleted.
The unit must be identical with the unit selected by
CALC:LIM:UNIT
. If no unit is indicated, the unit
defined with
CALC:LIM:UNIT
is automatically used.
In analyzer mode, the limit check signals errors if the measured values exceed the UPPer limit line.
Example:
"CALC:LIM2:UPP -10,0,0,-10,-5"
'Defines 5 upper limit values for limit line 2 in the preset unit.
"CALC:LIM2:UPP?"
'Outputs the upper limit values for limit line 2 separated by a comma.
Characteristics:
*RST value: - (LIMit:STATe is set to OFF)
SCPI: conform
Mode:
R, A
CALCulate<1|2>:LIMit<1...8>:UPPer:STATe
ON | OFF
This command switches on or off the indicated limit line. The limit check is activated separately with
CALC:LIM:STAT ON.
Up to 8 limit lines can be defined at the same time. The numeric suffixes <1|2> of CALCulate indicate
the measurement window.
In analyzer mode, the result of the l imit check can be queried with
CALCulate:LIMit<1...8>:
FAIL?
.
Example:
"CALC1:LIM4:UPP:STAT ON"
'Switches on limit line 4 (upper limit) in screen A.
"CALC2:LIM4:UPP:STAT ON"
'Switches on limit line 4 (upper limit) in screen B.
Characteristics:
*RST value: OFF
SCPI: conform
Mode:
R, A
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...