4.121
R&S ESU
Instrument Functions
Analyzer Mode
SINGLE
SWEEP
The
SINGLE SWEEP
softkey starts n sweeps after triggering. The number of
sweeps is determined by the
SWEEP COUNT
softkey.
When working in the split-screen mode, the frequency ranges of the two windows
are swept one after the other.
If a trace is swept using
TRACE AVERAGE
or
MAXHOLD,
the value set via the
SWEEP COUNT
softkey determines the number of sweeps. If 0 has been entered,
one sweep is performed.
Remote command:
INIT:CONT OFF
CONTINUE
SGL SWEEP
The
CONTINUE SGL SWEEP
softkey repeats the number of sweeps set under
SWEEP COUNT
, however without first deleting the trace.
This is particularly of interest when using the functions
TRACE AVERAGE
and
MAXHOLD
, if previously recorded measurement results are to be taken into consid-
eration for averaging / maximum search.
If
SGL SWEEP DISP OFF
is active, the screen is switched off also during repeated
sweeps.
Remote command:
INIT:CONM
SWEEPTIME
MANUAL
The
SWEEPTIME MANUAL
softkey activates the window for entering the sweep
time manually (see also
BW
menu).
Remote command:
SWE:TIME 10s
SWEEPTIME
AUTO
The
SWEEPTIME AUTO
softkey activates the automatic selection of the sweep time
as a function of the bandwidth of the resolution and video filters (see also
BW
menu).
Remote command:
SWE:TIME:AUTO ON
SWEEP
COUNT
The
SWEEP COUNT
softkey activates the window for the entry of the number of
sweeps to be performed by R&S ESU after a single sweep has been started. If
Trace Average, Max Hold or Min Hold is activated, this also determines the number
of averaging or maximum search procedures.
Example
[
TRACE1
:
MAX HOLD
]
[
SWEEP
:
SWEEP COUNT
: {10}
ENTER
]
[
SINGLE SWEEP
]
R&S ESU performs the Max Hold function over 10 sweeps.
The permissible range for the sweep count is 0 to 32767. For sweep count = 0 or 1,
one sweep is performed. For trace averaging in the continuous-sweep mode,
R&S ESU performs running averaging over 10 sweeps if sweep count = 0; if sweep
count = 1, no averaging, maxhold or minhold is performed.
The sweep count is valid for all the traces in a diagram.
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...