                Pulse Oximeter 

   

JPD-500E(LED) 

Issuing Date: April, 2020   

 

Manual Version: 1.0 

  Instructions to User 

 
Dear users, thank you very much for purchasing the Pulse Oximeter. 
This Manual is written and compiled in accordance with the council directive 
MDD 93/42/EEC for medical devices and harmonized standards. In case of 
modifications and software upgrades, the information contained in this 
document is subject to change without notice. 
Please read the User Manual carefully before using this product. 
 

WARNING: 

■    Uncomfortable or painful feeling may appear if using the device 
ceaselessly, especially for the microcirculation barrier patient. It is 
recommended that the sensor should not be applied to the same finger for 
over 2 hours. 

■    The light (the infrared is invisible) emitted from the device is harmful to the 
eyes, so the user and the maintenance man should not stare at the light. 

■    Enamel or Acrylic fingernail polish or other fingernail applications may 
distort and/or produce inaccurate readings. 

■    Please refer to the correlative literature about the clinical restrictions and 
caution. 

■    This device is not intended for treatment. 
 

  1 Safety 

 

1.1 Instructions for Safe Operations 

■    Check the main unit and all accessories periodically to make sure that 
there is no visible damage that may affect patient

’s safety and monitoring 

performance about cable and transducers. It is recommended that the device 
should be inspected once a week at least. When there is obvious damage, 
stop using the monitor. 

■    Necessary maintenance must be performed by qualified service 
engineers ONLY. Users are not permitted to maintain it by themselves. 

■    The oximeter cannot be used together with devices not specified in User 
Manual. Only the accessory that appointed or recommendatory by 
manufacture can be used with this device. 

■    This product is calibrated before leaving factory. 

1.2 Warning 

■    Explosive hazard---DO NOT use the oximeter in environment with 
inflammable gas such as some ignitable anesthetic agents. 

■    DO NOT use the oximeter while the user is measured by MRI and CT. 

■    The person who is allergic to rubber can not use this device. 

■    The disposal of scrap instrument and its accessories and packing 
(including battery, plastic bags, foams and paper boxes) should follow the 
local laws and regulations. 

■    Please check the packing before use to make sure the device and 
accessories are totally in accordance with the packing list, or else the device 
may have the possibility of working abnormally. 

1.3 Attentions 

●    Keep the oximeter away from dust, vibration, corrosive substances, 
explosive materials, high temperature and moisture. 

●    If the oximeter gets wet, please stop operating it. 

●    When it is carried from cold environment to warm or humid environment, 
please do not use it immediately. 

●    High temperature or high pressure steam disinfection of the oximeter is 
not permitted. Refer to User Manual in the relative chapter for instructions of 
cleaning and disinfection. 

●    Do not have the oximeter immerged in liquid. When it needs cleaning, 
please wipe its surface with medical alcohol by soft material. Do not spray 
any liquid on the device directly. 

●    Do not use the device on infant or neonatal patients. 

●    The product is suitable for children above four years old and 
adults(Weight should be between 15kg to 110kg). 

Summary of Contents for JPD-500E

Page 1: ...lified service engineers ONLY Users are not permitted to maintain it by themselves The oximeter cannot be used together with devices not specified in User Manual Only the accessory that appointed or recommendatory by manufacture can be used with this device This product is calibrated before leaving factory 1 2 Warning Explosive hazard DO NOT use the oximeter in environment with inflammable gas suc...

Page 2: ...eature Operation of the product is simple and convenient The product is small in volume light in weight total weight is about 50g including batteries and convenient in carrying Power consumption of the product is low The product will automatically be powered off when no signal is in the product within 16 seconds 2 3 Major Applications and Scope of Application The Pulse Oximeter can be used to meas...

Page 3: ...fluorescent lamp dual ruby light infrared heater direct sunlight and etc 4 Strenuous action of the subject or extreme electrosurgical interference may also affect the accuracy 4 Accessories 1 One hanging rope 2 Storage bag 2 Two AAA batteries 3 One user s manual 5 Installation 5 1 View of the Front panel Figure 2 Front View 5 2 Battery Step 1 Refer to Figure 3 and insert the two AAA size batteries...

Page 4: ... air or clean it by dry clean fabric Using the medical alcohol to disinfect the product after use prevent from cross infection for next time use Please take out the batteries if the oximeter is not use for a long time Warning High pressure sterilization cannot be used on the device Warning Do not immerse the device in liquid Warning It is recommended that the device should be kept in a dry environ...

Page 5: ...ormation Display Mode Display Format LED display The Pulse Oxygen Saturation SpO2 Digital Pulse Rate PR Digital Pulse Intensity bar graph Digital bar graph display SpO2 Parameter Specification Measuring range 35 100 the resolution is 1 Accuracy 70 100 2 Below 70 unspecified Pulse Parameter Specification Measuring range 25bpm 250bpm the resolution is 1 bpm Accuracy 2bpm Pulse Intensity Range Contin...

Page 6: ...dical com Postal Code 518103 DE Überblick Die Sauerstoffsä ttigung ist der Prozentsatz von Oxyhä moglobin HbO2 der mit Sauerstoff gegen alle kombinierbaren Hä moglobinen Hb kombiniert wird Es ist ein wichtiger physiologischer Parameter der in die Atmung und den Kreislauf einbezogen ist Mit diesem Instrument kann die Pulsfrequenz über die Arterie gemessen werden Produktmerkmale Leicht tragbar und e...

Page 7: ... von der Bildschirmanzeige 5 Drücken Sie im Boot Strap Zustand die Taste und das Gerä t wird zurückgesetzt Fingernä gel und Lumineszenzrohr sollten auf derselben Seite liegen Fehlersymptome Ursachen und Lösungen Symptom Möglice Ursache Solution Das Oximeter zeigt keine Daten zur Sauerstoffsä ttigung und Pulsfrequen z an 1 Der Finger befindet sich nicht in der richtigen Position 2 Der Patient konze...

Page 8: ...it Poids léger utilisation portable et opération simple Écran LED lisible Indication de basse tension de la batterie L appareil s éteindra automatiquement 16 secondes plus tard si aucun signal n est détecté L appareil est équipé de deux piles sèches 1 5 V AAA qui peuvent lui fournir de l énergie pour fonctionner plus de 24 heures vue du panneau avant 1 affichage basse tension 2 fréquence du pouls ...

Page 9: ...tacter le centre de service local L écran s éteint soudainement 1 L appareil s éteint automatiquement en l absence de signal dans les 16 secondes 2 Les batteries sont presque épuisées 1 Normal 2 Changez les piles IT Panoramica La saturazione di ossigeno è la percentuale di ossiemoglobina HbO2 che è combinata con l ossigeno contro tutta l emoglobina combinabile Hb È un parametro fisiologico importa...

Page 10: ...l polso non sono visualizzati stabilmente 1 Il dito non è inserito abbastanza in profondità 2 Il dito trema o il paziente si sta muovendo 1 Posizionare il dito correttamente e riprovare 2 Lascia che il paziente mantenga la calma Il dispositivo non può essere acceso 1 Batteria scarica o nessuna batteria 2 Le batterie non sono inserite correttamente 3 Il malfunzionamento del dispositivo 1 Sostituire...

Page 11: ...lema Posible razón Solución La SpO2 y la frecuencia del pulso no se muestra normalmente 1 El dedo no estácolocado correctamente 2 El valor de la SpO2 en el paciente es demasiado baja para ser detectado 1 Coloque el dedo correctamente y vuelva a intentarlo 2 Inténtelo de nuevo Vaya al hospital para un diagnó stico si está seguro de que el dispositivo funciona correctamente La SpO2 y la frecuencia d...

Page 12: ...treet London W1G 9QR UK Shenzhen Jumper Medical Equipment Co Ltd Address D Building No 71 Xintian Road Fuyong Street Baoan Shenzhen Guangdong China E mail info jumper medical com Tel 86 755 26692192 26696279 Web www jumper medical com ...

Reviews: