Operating Manual 1302.6163.12 - 03
4.249
R&S ESU
Instrument Functions
Basic Settings
The preset value is at -200 dBm. The field is displayed if the value RELATIVE is
entered in the field Y-SCALING.
Remote command:
CALC:LIM3:UPP:THR -30 dBm
or
CALC:LIM3:LOW:THR -30 dBm
Comment
- Enter comments
Comments are arbitrary, however, they must be less than 41 characters long.
Remote command:
CALC:LIM3:COMM "Upper limit"
VALUES
The
VALUES
softkey activates the entry of the data points in the table columns
Time
/
Frequency
and
Limit/dB
. Which table columns appear depends upon the
Domain
selection in the table heading.
The desired frequency/time data points are entered in ascending order (two
repeated frequencies/time values are permitted).
Remote command:
CALC:LIM3:CONT:DATA 1MHz,3MHz,30MHz
CALC:LIM3:UPP:DATA -10,0,0
CALC:LIM3:LOW:DATA -30,-40,-40
INSERT VALUE
The
INSERT VALUE
softkey creates an empty line above the current cursor position
where a new data point may be entered. However, during the entry of new values, it
is necessary to observe an ascending order for frequency/time.
Remote command:
--
DELETE VALUE
The
DELETE VALUE
softkey erases the data point (complete line) at the cursor
position. All succeeding data points are shifted down accordingly.
Remote command:
--
SHIFT X LIMIT
LINE
The
SHIFT X LIMIT LINE
softkey calls an entry window where the complete limit line
may be shifted parallel in the horizontal direction.
The shift takes place according to the horizontal scale:
– in the frequency domain in Hz, kHz, MHz or GHz
– in the time domain in ns, µs, ms or s
In this manner, a new limit line can be easily generated based upon an existing limit
line which has been shifted horizontally and stored (
SAVE LIMIT LINE
softkey)
under a new name (
NAME
softkey).
Remote command:
CALC:LIM3:CONT:SHIF 50KHz
SHIFT Y LIMIT
LINE
The
SHIFT Y LIMIT LINE
softkey calls an entry window where the complete limit line
may be shifted parallel in the vertical direction.
The shift takes place according to the vertical scale:
– for logarithmic units, relative, in dB
– for linear units, as a factor
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...