R&S ESU
Instrument Functions
Basic Settings
4.282
Operating Manual 1302.6163.12 - 03
This softkey is only available, if a HP languge is selected via the
softkey.
Remote command:
SYST:RSW ON
COUPLING
FSP/HP
The
COUPLING FSP/HP
softkey controls the default coupling ratios for:
• span and resolution bandwidth (Span/RBW) and
• resolution bandwidth and video bandwidth (RBW/VBW)
for the HP emulation mode.
In case of FSP the standard parameter coupling of the instrument is used. As a
result in most cases a shorter sweeptime is used than in case of HP.
The softkey is only available in HP emulation mode.
Remote command:
SYST:HPC FSP
User Port Configuration
The instrument provides a parallel interface, which is 8 bits wide. Over these ports,
arbitrary bit patterns can be output or input.
USER PORT
The
USER PORT
softkey opens the submenu for setting the direction of the data
transmission and for entering the values of the individual ports.
USER PORT IN/
OUT
The USER PORT IN/OUT determines in which direction the interface transmits data.
USER PORT (read operation)
The value is indicated in a window. A new readout is started by pressing
READ
.
Pressing
OK
closes the window
Remote command:
INP:UPOR:STAT ON
INP:UPOR?
USER PORT OUT (write operation)
The bit pattern selected with softkeys
PORT 0
to
PORT 7
is output.
Remote command:
OUTP:UPOR:STAT ON
PORT 0 0/1 to
PORT 7 0/1
The
PORT 0 0/1
to
PORT 7 0/1
softkeys determine the output value for port 1 to 7.
These softkeys are only available for setting
USER PORT OUT
.
Remote command:
OUTP:UPOR STAT ON
OUTP:UPOR #B10110010
00000000
USER PORT INPUT
READ
OK
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...