Operating Manual 1302.6163.12 - 03
6.207
R&S ESU
Remote Control – Description of Commands
SENSe Subsystem
6.16.12
SENSe:ROSCillator Subsystem
This subsystem controls the reference oscillator. The numeric suffix in SENSe is irrelevant for the com-
mands of this subsystem.
[SENSe<1|2>:]ROSCillator:SOURce
INTernal | EXTernal | EAUTo
This command controls selection of the reference oscillator.
If the external reference oscillator is selected, the reference signal must be connected to the rear panel
of the instrument.
As of firmware version 4.4x, you can you also select EAUTo. This paramater selects an external
reference signal, but switches to the internal reference signal if no external reference is detected. If the
external reference oscillator is selected (EXT | EAUT), the reference signal must be connected to the
rear panel of the instrument.
The reference settings are not changed after a preset.
Example:
"ROSC:SOUR EXT"
Characteristics:
*RST value: –
SCPI: conform
Mode:
R, A
*RST
is not
influencing this parameter.
[SENSe<1|2>:]ROSCillator:SOURce:EAUTo?
This command queries the current source of the reference signal, if the external reference source EXT
(INT) is used. The command returns INT if the internal reference source is used and EXT for an
external reference source.
This command is a query and therefore has no *RST value.
Example:
"ROSC:SOUR:EAUT?"
Characteristics:
*RST value: -
SCPI: device-specific
Mode:
R, A
INTernal
Uses the internal reference signal.
EXTernal
Uses the external reference signal.
EAUTo
Uses the external reference signal, but automatically switches to the internal
reference signal, if no external reference signal is detected. Note that you have
to change the reference source back to EXT and EAUT once the connection to
the external reference is reestablished.
You can query the current status with
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...