Remote Control – Description of Commands
SENSe Subsystem
[SENSe<1|2>:]POWer:ACHannel:PRESet:RLEVel:OFFSet
0 … 100 dB
This command allows to modify the ADJUST REF LEVEL function for user standards. It defines an
additional level offset to be taken into account for the analyzer's reference level setting. It is only
available if Fast ACP is not active and no predefine standard is selected.
Example:
"POW:ACH:PRES:RLEV:OFFS 10dB"
'set the level adjust offset to 10dB
Characteristics:
*RST value: 0 dB
SCPI: device-specific
Mode:
A
[SENSe<1|2>:]POWer:ACHannel:REFerence:AUTO
ONCE
This command sets the reference value for the relative measurement to the currently measured
channel power.
The command is only available in the frequency domain (span > 0).
Example:
"POW:ACH:REF:AUTO ONCE"
Characteristics:
*RST value: -
SCPI: device-specific
Mode:
A-F
This command is an event and therefore has no *RST value and no query.
[SENSe<1|2>:]POWer:ACHannel:REFerence:TXCHannel:AUTO
MINimum | MAXimum | LHIGhest |
OFF
This command activates or deactivates the automatic selection of a transmission channel to be used
as a reference channel in relative adjacent channel power measurements.
The transmission channel with the highest power, the transmission channel with the lowest power,
or the transmission channel nearest to the adjacent channels can be defined as a reference channel.
The command is available only for multi carrier channel and adjacent channel power measurements
(
CALC:MARK:FUNC:POW:SEL MCAC
) in the frequency domain (span > 0).
Parameter:
MINimum: Transmission channel with the lowest power
MAXimum: Transmission channel with the highest power
LHIGhest: Lowermost transmission channel for the lower adjacent channels,
uppermost transmission channel for the upper adjacent channels
OFF: Deactivates the automatic transmission channel selection.
Example:
"POW:ACH:REF:TXCH:AUTO MAX"
'The transmission channel with the highest power is used as a reference channel.
Characteristics:
*RST value: -
SCPI: device-specific
Mode:
A-F
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...