Operating Manual 1302.6163.12 - 03
4.323
R&S ESU
Instrument Functions
Tracking Generator – Option R&S FSU-B9
4.10.2.2
Normalization
NORMALIZE
The
NORMALIZE
softkey switches the normalization on or off. The softkey is only
available if the memory contains a correction trace.
It is possible to shift the relative reference point within the grid using the
REF VALUE
POSITION
softkey. Thus, the trace can be shifted from the upper border of the grid
to the vertical center of the grid:
Fig. 4.22 Normalized display
In
SPLIT SCREEN
operation, the normalization is switched on in the currently active
window. Different types of normalization can be active in the two windows.
Normalization is aborted when the
NETWORK
operating mode is quit.
Remote command:
CORR ON
REF VALUE
POSITION
The
REF VALUE POSITION
softkey marks a reference position in the active window
at which the normalization result (calculated difference with a reference trace) is dis-
played.
If no reference line is active, the softkey switches on a reference line and activates
the input of its position. The line can be moved within the grid boundaries.
The reference line is switched off by pressing the softkey again.
The function of the reference line is explained in the section
.
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...