R&S ESU
Remote Control – Description of Commands
SENSe Subsystem
6.176
Operating Manual 1302.6163.12 - 03
[SENSe<1|2>:]ESPectrum:RTYPe
PEAK | CPOWer
This command sets selects the type of power measurement in the reference range:
The numeric suffixes <1...20> specify the range. The numeric suffix at SENSe<1|2> is irrelevant.
Parameter:
PEAK: Measures the highest peak within the reference range.
CPOWer: Measures the channel power within the reference range (integral
bandwidth method).
Example:
"ESP:RTYP PEAK"
'selects highest peak
Characteristics:
*RST value: CPOW
SCPI: device-specific
Mode:
A