R&S ESU
Remote Control – Description of Commands
INITiate Subsystem
6.128
Operating Manual 1302.6163.12 - 03
INITiate<1|2>:EMITest
In receiver mode, this command starts an automatic sequence consisting of the preliminary
measurement, determination of the peak list and subsequent final measurement.
Example:
"INIT2:EMIT"
'Starts the sequence preliminary measurement, peak search and final
measurement
Characteristics:
*RST value: -
SCPI: device-specific
Mode:
R
This command is an event, so it has neither an *RST value nor a query function.
INITiate<1|2>:SPURious
This command starts a new spurious measurement.
Example:
"INIT:CONT OFF"
'switches to single-sweep mode
"INIT:SPUR;*WAI"
'starts the measurement by waiting for the end of the 20 measurements
Characteristics:
*RST value: -
SCPI: device-specific
Mode:
A
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...