Operating Manual 1302.6163.12 - 03
6.163
R&S ESU
Remote Control – Description of Commands
SENSe Subsystem
[SENSe<1|2>:]CORRection:TRANsducer:VIEW
ON | OFF
This command switches on the display of the active transducer factor or set. Prior to this command,
the command
SENS:CORR:TRAN:SEL
must be sent.
This command is available from firmware version 1.40 or higher.
Example:
"CORR:TRAN:SEL 'FACTOR1'"
"CORR:TRAN:VIEW ON"
Characteristics:
*RST value: OFF
SCPI: device-specific
Mode:
R, A
[SENSe<1|2>:]CORRection:TRANsducer:ADJust:RLEVel[:STATe]
ON | OFF
This command switches the automatic adjustment of the reference level to the selected transducer
factor on or off. Prior to this command, the command
SENS:CORR:TRAN:SEL
must be sent.
Example:
"CORR:TRAN:SEL 'FACTOR1'"
"CORR"CORR:TRAN:ADJ:RLEV ON"
Characteristics:
*RST value: OFF
SCPI: device-specific
Mode:
R, A
[SENSe<1|2>:]CORRection:TSET:SELect
<name>
This command selects the transducer set designated by <name>. If <name> does not exist yet, a new
set is created.
This command must be sent prior to the subsequent commands for modifying/activating
transducersets.
Parameter:
<name>::= name of the transducer set in string data form with a maximum of 8
characters.
Example:
"CORR:TSET:SEL 'SET1'"
Characteristics:
*RST value: -
SCPI: device-specific
Mode:
R, A
[SENSe<1|2>:]CORRection:TSET:UNIT
<string>
This command defines the unit of the selected transducer sets. When assigning transducer factors to
the set, only factors which are compatible to the selected unit, i.e. factors with the same unit or the unit
dB, are allowed. Prior to this command, the command
SENS:CORR:TSET:SEL
must be sent.
Parameter:
<string>::= 'DB' | 'DBM' | 'DBUV' | 'DBUV/M' | 'DBUA' | 'DBUA/M'' | DBPW' | 'DBPT'
Example:
"CORR:TSET:SEL 'SET1'"
"CORR:TSET:UNIT 'DBUV'"
Characteristics:
*RST value: 'DB'
SCPI: device-specific
Mode:
R, A
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...