Operating Manual 1302.6163.12 - 03
4.271
R&S ESU
Instrument Functions
Basic Settings
TRANSDUCER
FACTOR
The
TRANSDUCER FACTOR
softkey places the scrollbar on the position of the
active transducer factor.
If a transducer factor is not active, the scrollbar is placed on the first line of the table.
Remote command:
CORR:TRAN:SEL <name>
CORR:TRAN ON | OFF
TRANSDUCER
SET
The
TRANSDUCER SET
softkey places the scrollbar on the position of the active
transducer set. For further details refer to
“Entry and Editing of Transducer Sets” on
If no transducer set is switched on, the bar is set to the first line of the table.
Remote command:
CORR:TSET:SEL <name>
CORR:TSET ON |OFF
EDIT
The
EDIT TRD FACTOR
softkey gives access to the submenu for editing and gener-
ating transducer factors. For details refer to section
“Entry and Editing of Transducer
.
NEW
The
NEW FACTOR
softkey gives access to the submenu for editing and generating
transducer factors. For details refer to section
“Entry and Editing of Transducer Fac-
DELETE
The
DELETE
softkey
deletes the marked factor or set.
To prevent deletion by mistake, deletion has to be confirmed.
Remote command:
CORR:TRAN DEL
VIEW
TRANSDUCER
The
VIEW TRANSDUCER
softkey activates the display of the active transducer fac-
tor or set. During the measurement, the interpolated characteristic of the active
transducer factor or set is shown in the diagram instead of the measured values.
Remote command:
CORR:TRAN:VIEW ON
VIEW
TRANSDUCER
The
VIEW TRANSDUCER
softkey activates the display of the active transducer fac-
tor or set. During the measurement, the interpolated characteristic of the active
transducer factor or set is shown in the diagram instead of the measured values.
Remote command:
CORR:TRAN:VIEW ON
REFLVL ADJ
AUTO MAN
When a transducer factor is used, the trace is moved by a calculated shift. However,
an upward shift reduces the dynamic range for the displayed values. With the soft-
key
REFLVL ADJ
an automatic reference level offset adaptation allows to restore
the original dynamic range by also shifting the reference level by the maximum
value of the transducer factor.
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...