Operating Manual 1302.6163.12 - 03
6.285
R&S ESU
Remote Control – Description of Commands
GPIB Commands of HP Models 856xE, 8566A/B, 8568A/B and 8594E
6.23.5
Special Behavior of Commands
Command
Known Differences
ABORT
Does not automatically set the command complete bit (bit 4) in the status byte. An additional DONE is
required for that purpose.
ANNOT
Only frequency axis annotation is affected.
AT
AT DN/UP: Step size
CAL
The CAL commands do not automatically set the command complete bit (bit 4) in the status byte. An
additional DONE command is required for that purpose.
CF
Default value, range, step size
CR
Default ratio Span / RBW
CT
Formula for coupled sweeptime
CV
Default ratio RBW / VBW
DET
DET? returns SAMP instead of SMP on the R&S ESU.
DET not automatically set the command complete bit (bit 4) in the status byte. An additional DONE is required
for that purpose.
ERR?
Deletes the error bit in the status register but always returns a '0' in response.
FA
Default value, range, step size
FB
Default value, range, step size
ID
Query of instrument type. The instrument type entered with
SETUP - GENERAL SETUP - GPIB - ID STRING
USER
will be returned.
M2
Default value, range, step size
M3
Default value, range, step size
MKACT
Only marker 1 is supported as the active marker.
MKBW
Default value
MKPT
Step size
MKPX
Step size
NRL
OL?
Storage of instrument settings:
80 characters are returned as information on the instrument settings.
The contents of the 80 characters returned does not correspond to the original data contents of the 8566A /
8568A family.
OL
Readout of instrument settings:
The 80 characters read by means of OL? are accepted as information on the corresponding instrument
settings.
The contents of the 80 characters read does not correspond to the original data contents of the 8566A /
8568A family.
RB
Default value, range, step size
RL
Default value, step size
RLPOS
Adapts the position of the reference level even if the tracking generator normalization is not active.
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...