R&S ESU
Remote Control – Description of Commands
Differences in GPIB Behavior between the FSP and the FSE Families of Instruments
6.290
Operating Manual 1302.6163.12 - 03
R&S
ESU
FSE
*TRG
R&S ESU starts
measurement in active
screen
FSE: starts
measurement in both
screens (split screen
mode)
R&S
ESU
FSE
*TST?
R&S
ESU
FSE
*WAI
R&S
ESU
FSE
ABORt
R&S
ESU
CALCulate:STATistics:APD[:STATe]
ON | OFF
new function for
R&S ESU
R&S
ESU
CALCulate:STATistics:CCDF[:STATe]
ON | OFF
new function for
R&S ESU
R&S
ESU
CALCulate:STATistics:NSAMples
100 to 1E9
new function for
R&S ESU
R&S
ESU
CALCulate:STATistics:PRESet
new function for
R&S ESU
R&S
ESU
CALCulate:STATistics:Result<1...3>?
MEAN | PEAK | CFACtor
| ALL
new function for
R&S ESU
R&S
ESU
CALCulate:STATistics:SCALe:AUTO
ONCE
new function for
R&S ESU
R&S
ESU
CALCulate:STATistics:X:RANGe
-10dB to 200dB
new function for
R&S ESU
R&S
ESU
CALCulate:STATistics:X:RLEVel
-130dBm to 30dBm
new function for
R&S ESU
R&S
ESU
CALCulate:STATistics:Y:LOWer
-1E-9 to 0.1
new function for
R&S ESU
R&S
ESU
CALCulate:STATistics:Y:UPPer
-1E-8 to 1.0
new function for
R&S ESU
FSE
CALCulate<1|2>:CTHReshold
MIN to MAX
not available in
R&S ESU
FSE
CALCulate<1|2>:CTHReshold:STATe
ON | OFF
not available in
R&S ESU
R&S
ESU
CALCulate<1|2>:DELTamarker<1...4>:AOFF
markers 2...4 are either
normal or delta markers;
marker 1 always serves
as the reference marker
for all deltamarkers
FSE
CALCulate<1|2>:DELTamarker<1...4>:AOFF
there are 4markers and
4 deltamarkers; the most
recently used marker
serves as the reference
marker for all
deltamarkers
Devices
Command
(Sheet 2 of 47)
Parameter
Notes
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...