Operating Manual 1302.6163.12 - 03
4.25
R&S ESU
Receiver Mode
4.5.4
Selection of the Measurement Function – MEAS
Key
The
MEAS
key opens the menu to select the detectors and set the measurement
time for the receiver and the audio demodulator.
To reduce the amount of data in RFI voltage measurements, a list of subrange max-
ima (softkey
PEAK SEARCH
) can be generated from the scan results and an accep-
tance line (softkey
MARGIN
) may be defined in the
FINAL MEAS
submenu. A final
measurement is performed only for frequencies with high RFI level.
The final measurement may be either automatic (softkey
AUTOMATIC FINAL
) or in
interactive mode (softkey
INTERACTIVE
). Automatic control functions for line
impedance stabilization networks (LISN) are available in the side menu of the
FINAL
MEAS
submenu. The final measurement is started with the softkey
RUN FINAL
MEAS
.
The
THRESHOLD SCAN
function is also used to reduce data. The scan is immedi-
ately interrupted when a limit violation is detected and the final measurement per-
formed within a short time.
MEAS
Softkeys in the
MEAS
menu:
!
!
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...