Operating Manual 1302.6163.12 - 03
4.31
R&S ESU
Receiver Mode
MEAS TIME
The
MEAS TIME
softkey activates the entry field for the measurement time.
This softkey is also available in the RECEIVER / MEAS menu.
Remote command:
SWE:TIME 100ms
Effect of measurement time with different weighting modes:
• MIN/MAX PEAK measurement:
With peak weighting selected, the maximum or minimum level during the selected
measurement time is displayed. The peak detector is reset at the beginning of
each measurement. At the end of the measurement time the maximum or
minimum level occurred during the measurement time is displayed. Since the
peak detector of R&S ESU is a digital detector, discharging is irrelevant even with
long measurement times.
Unmodulated signals can be measured with the shortest possible measurement
time. When pulses are measured, the selected measurement time must be long
enough for at least one pulse to occur during the measurement time.
• Average measurement:
With average detection selected, the video voltage (envelope of IF signal) is
averaged during the measurement time. Averaging is digital, i.e. the digitized
values of the video voltage are summed up and divided by the number of samples
at the end of the measurement time. This corresponds to a filtering with a
rectangular window in the time domain and a filtering with sin x/x characteristic in
the frequency domain. With unmodulated signals the shortest possible
measurement time can be selected. With modulated signals the measurement
time is determined by the lowest modulation frequency to be averaged. With pulse
signals, the selected measurement time should be long enough for sufficient
number of pulses (>10) to occur in the measurement window for averaging.
Bandwidth
Shortest measurement time
AV, RMS
Shortest measurement time
PK+, PK-
≤
10 Hz
1 sec
10 msec
100 Hz
100 msec
1 msec
200 Hz ... 300 Hz
50 msec
1 msec
1 kHz ... 3 KHz
10 msec
0,1 msec
9 kHz ... 30 kHz
1 msec
0,1 msec
≥
100 kHz
0,1 msec
5 µsec
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...