R&S ESU
Instrument Functions
Analyzer Mode
4.194
Operating Manual 1302.6163.12 - 03
➢
Create a new user standards with
➢
Select the user standard you just created with
➢
Change the offset
➢
Save the user standard again to keep the current settings.
Remote command:
POW:ACH:PRES:RLEV:OFFS 10dB
SELECT TRACE
The
SELECT TRACE
softkey selects the trace on which the CP/ACP measurement
is to be performed. Only activated traces can be selected, i.e. traces not set to
BLANK.
Remote command:
SENS:POW:TRAC 1
Examples
Measurement of adjacent-channel power for a specific standard:
The adjacent-channel power is to be measured for a signal at 800 MHz with 0 dBm
level in line with IS136.
[
PRESET
]
Set the R&S ESU to the default setting.
[
FREQ
:
CENTER
: 800
MHz
]
Set the center frequency to 800 MHz.
[
AMPT
: 0
dBm
]
Set the reference level to 0 dBm.
[
MEAS
]
Call the menu for the measurement functions.
[
CHAN PWR / ACP
]
Select the channel and adjacent-channel power measure-
ment function. The measurement is performed with the
default settings or a previously defined setting. The sub-
menu for setting the desired new configuration is opened.
[
CP/ACP STANDARD
:
select IS136:
ENTER
]
Select the NADC (IS136) standard.
[
CP/ACP CONFIG
]
Call the submenu for configuration of the adjacent-channel
power measurement.
[
NO. OF ADJ CHAN
: 2
ENTER
]
Select two adjacent channels for the measurement, i.e.
the adjacent channel and the alternate channel are mea-
sured.
[
ADJUST SETTINGS
]
Set the optimum span, resolution bandwidth (RBW), video
bandwidth (VBW) and detector automatically for the mea-
surement. The absolute channel power and the relative
power of the adjacent channels are displayed on the
screen.
Change to the main menu for channel power measure-
ment.
[
ADJUST REF LVL
]
Set the reference level equal to the channel power mea-
sured.
PREV
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...