R&S ESU
Remote Control – Description of Commands
CALCulate Subsystem
6.86
Operating Manual 1302.6163.12 - 03
TRACE 1 of the selected screen is always used by the function. The suffix of MARKer will be ignored.
Example:
"DISP:WIND:TRAC:Y:RLEV –10dBm"
'Sets the reference level to 10 dBm
"INP:ATT 30 dB"
'Sets the input attenuation to 30 dB
"FREQ:CENT 935.2MHz;SPAN 0Hz"
'Sets the receive frequency to 935.2 MHz and the span to 0 Hz
"BAND:RES 1MHz;VID 3MHz"
'Sets the resolution bandwidth to 1 MHz and the video bandwidth to 3 MHz
"DET RMS"
'Sets the RMS detector
"TRIG:SOUR VID;LEV:VID 50 PCT"
'Selects the trigger source VIDeo and sets the level of the video trigger source to
50 PCT
"SWE:TIME 50ms"
'Sets the sweep time to 50 ms
"INIT;*WAI"
'Starts the measurement with synchronization
"CALC:MARK:FUNC:MSUM? 50US,450US,576.9US,8"
'Queries 8 bursts with an offset of 50 µs, a test time of 450 µs and a period of
576.9 µs
Characteristics:
*RST value: -
SCPI: device-specific
Mode:
A-T
t
Trace start
Time offset of
first pulse
Measurement
Time
Measurement
Time
Measurement
Time
Period
Period
P
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...