6.5.2
CALCulate:ESPectrum Subsystem
The following chapter describes remote control commands that configure the list evaluation for the Spec-
trum Emission Mask measurement.
CALCulate<1|2>:ESPectrum:PSEarch | PEAKsearch:AUTO
ON | OFF
This command activates or deactivates the list evaluation of the Spectrum Emission Mask
measurement.
Example:
“
CALC:ESP:PSE:AUTO OFF
”
’Deactivates the list evaluation
Characteristics:
RST value: ON
SCPI: device-specific
Mode:
A
:CALCulate<1|2>:ESPectrum:PSEarch | PEAKsearch:MARGin
-200dB … +200dB
This command sets the margin used for the limit check/peak search of the Spectrum Emission Mask
measurement.
Example:
“CALC:ESP:PSE:MARG 10"
' sets the margin to 10 dB
Characteristics:
*RST value: 6 dB
SCPI: device-specific
Mode:
A
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...