INSTrument:COUPle
NONE | RLEVel | CF_B | CF_A
In analyzer mode, this command selects the parameter coupling between the two measurement
windows screen A and B.
Parameter:
NONE: No coupling. The two measurement windows are operated like two
independent "virtual" devices.
RLEVel: The reference levels of the two measurement windows are coupled.
CF_B: The center frequency of screen B is coupled to the frequency of marker 1 in
screen A.
CF_A: The center frequency of screen A is coupled to the frequency of marker 1 in
screen B.
Example:
"INST:COUP NONE"
'Switches off the coupling of measurement windows. This leads to two independent
"virtual" devices.
Characteristics:
*RST value: NONE
SCPI: device-specific
Mode:
A
INSTrument:COUPle:ATTenuation
ALL | NONE
This command couples die RF attenuation settings for the analyzer and receiver modes with one
another.
The RF attenuation unit and level unit are then set to the same value in the analyzer and receiver
screen.
Example:
"INST:COUP:ATT ALL"
Characteristics:
*RST value:ALL
SCPI:device-specific
Mode:
R, A
INSTrument:COUPle:BANDwidth|BWIDth
ALL | NONE
This command couples the bandwidth settings for the analyzer and receiver modes with one another.
The resolution bandwidth (RBW) and filter type are then set to the same value on the analyzer and
receiver screen. In some cases, other settings (e.g. the quasipeak detector) can block these settings.
Example:
"INST:COUP:BWID ALL"
Characteristics:
*RST value:NONE
SCPI:device-specific
Mode:
R, A
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...