R&S ESU
Remote Control – Description of Commands
CALCulate Subsystem
6.42
Operating Manual 1302.6163.12 - 03
6.5.4
CALCulate:MARKer Subsystem
The CALCulate:MARKer subsystem checks the marker functions in the instrument.
The measurement windows are selected via CALCulate1 (screen A) or 2 (screen B).
In receiver mode, the markers can only be activated after a scan has been carried out.
CALCulate<1|2>:MARKer<1...4>[:STATe]
ON | OFF
This command switches on or off the selected marker. If no indication is made, marker 1 is selected
automatically. If marker 2, 3 or 4 is selected and used as a delta marker, it is switched to marker mode.
The measurement windows are selected via CALCulate1 (screen A) or 2 (screen B).
Example:
"CALC:MARK3 ON"
'Switches marker 3 in screen A on or to marker mode.
Characteristics:
*RST value: OFF
SCPI: device-specific
Mode:
R, A
CALCulate<1|2>:MARKer<1...4>:AOFF
This command switches off all active markers in the indicated measurement window.
All delta markers and active marker/delta marker measurement functions are also switched off.
Example:
"CALC:MARK:AOFF"
'Switches off all markers in screen A.
Characteristics:
*RST value: -
SCPI: device-specific
Mode:
R, A
This command is an event and therefore has no *RST value and no query.
CALCulate<1|2>:MARKer<1...4>:TRACe
1 to 3
This command assigns the selected marker (1...4) to the indicated measurement curve. The
corresponding trace must be active, i.e. its status must be different from "BLANK".
If necessary, the corresponding marker is switched on prior to the assignment.
The measurement windows are selected via CALCulate1 (screen A) or 2 (screen B).
Example:
"CALC:MARK3:TRAC 2"
'Assigns marker 3 in screen A to trace 2.
"CALC2:MARK:TRAC 3"
'Assigns marker 1 in screen B to trace 3.
Characteristics:
*RST value: -
SCPI: device-specific
Mode:
R, A
CALCulate<1|2>:MARKer<1...4>:X
0 to MAX (frequency | sweep time)
This command positions the selected marker to the indicated :
• frequency (span > 0)
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...