
4. Operating Instruction
4.1 Function operation guide
82
4.1.3.1 Trace detector and trace operation
The trace refers to a set of sampling points of data, each of which (x, y) has one x value generally representing
frequency or time and y value generally representing amplitude. Each data point is also called trace point. In
any trace, the first data point is called trace 0, while the last one is called trace (sweep point -1). The trace
points are also called sweep points. For the spectrum trace, one trace point is generally corresponding to one
frequency interval, also called Bucket sometimes, and the measured value Y represents a measured data
corresponding to one Bucket. The menu corresponding to the
【
Trace
】
key on the front panel can help you
configure the measured trace and complete the trace operation.
Each trace can be considered to be one analysis implemented to the measured data. Each trace window of the
4051 Series Signal/Spectrum Analyzer can display 6 traces. The trace configurations determine how the
measured data analyzes and displays. The section mainly describes how to configure the trace and execute
trace operation.
1)
Concept of trace detector
When the sweep type is sweep under the spectrum analysis mode, the Series Signal Analyzer control the LO
by digital mode to carry out step sweep in smallest frequency. During the entire sweep period, the sampling
points captured by ADC in the digital IF circuit are much larger than the sweep points that used by the Series
Signal Analyzer for display. When the sweep type is FFT, the Series Signal Analyzer tunes the LO to fixed
frequency point. During the entire measurement period, after the FFT is carried out, the sampling points
captured by ADC in the digital IF circuit may be larger than the sweep points that used by the Series Signal
Analyzer for display.
E.g., we assume that the sampling frequency of ADC is 100 msa/s, sweep points are 1000, sweep time is 100
ms and the span is 1 GHz. For each sweep, the Series Signal Analyzer captures 100*10
5
sampling points,
which are needed to be turned into 1001 trace data. Equivalently, every 100,000 sampling data is
corresponding to the trace data at one point, while each trace datum represents the signals appeared with
frequency range of 1 MHz. If the sweep points are added, the span corresponding to each trace datum is
narrower, the measured results are more stable and the frequency readout accuracy is higher.
From the above analysis, it can be seen that the trace display needs to turn the more sampling data points into
the less sweep trace points, which is what the trace detector needs to do.
There are many types for the trace detector modes. The 4051 Series Signal/Spectrum Analyzer has 7 trace
detector modes:
Normal detection
Take the maximum value and minimum value from the sampling data assigned to each trace point and display
them at the same time.
Positive peak detection
Take and display the maximum value from the sampling data assigned to each trace point.
Negative peak detection
Take and display the minimum value from the sampling data assigned to each trace point.
Sampling Value Detector
Take and display the last value from the sampling data assigned to each trace point.
Power Average Detector
Also called root mean square (RMS) detector. Calculate the root mean square of all sampling data assigned to
each trace point; the 4051Series Signal Analyzer envelopes the detector for the RF input signal and obtains
linear voltage. After square these voltage values, take sum and divide the count of sampling data points
corresponding to each trace point. Finally, take square root. When the scale is logarithm, take 20 times of
logarithm transformation for these RMS values (with 10 as base) and obtain the trace data. When the scale
type is linear, those RMS values are trace data.
Summary of Contents for 4051 Series
Page 5: ......
Page 6: ......
Page 7: ......
Page 14: ......
Page 18: ......
Page 68: ......
Page 225: ...5 Menu 5 1 Menu structure 211 5 1 3 13 Maker Mkr Fig 5 3 13 Marker Menu...
Page 241: ...5 Menu 5 1 Menu structure 227 5 1 4 13 Maker Mkr Fig 5 4 13 Marker Menu...
Page 259: ...5 Menu 5 1 Menu structure 245 5 1 5 13 Maker Mkr Fig 5 5 13 Marker Menu...
Page 266: ...5 Menu 5 1 Menu structure 252 5 1 6 5 Sweep Sweep Fig 5 6 5 Sweep Menu...
Page 275: ...5 Menu 5 1 Menu structure 261 5 1 6 13 Maker Mkr Fig 5 6 13 Marker Menu...
Page 277: ...5 Menu 5 1 Menu structure 263 5 1 6 15 Peak Search Peak Search Fig 5 6 15 Peak Search Menu...
Page 280: ...5 Menu 5 1 Menu structure 266 5 1 7 3 Span X Scale Span Fig 5 7 3 Span Menu...
Page 291: ...5 Menu 5 1 Menu structure 277 5 1 7 13 Maker Mkr Fig 5 7 13 Marker Menu...
Page 296: ...5 Menu 5 1 Menu structure 282 5 1 8 3 Span X Scale Span Fig 5 8 3 Span Menu...
Page 306: ...5 Menu 5 1 Menu structure 292 5 1 8 12 Marker Marker Fig 5 8 12 Marker Menu...
Page 307: ...5 Menu 5 1 Menu structure 293 5 1 8 13 Maker Mkr Fig 5 8 13 Marker Menu...
Page 309: ...5 Menu 5 1 Menu structure 295 5 1 8 15 Peak Search Peak Search Fig 5 8 15 Peak Search Menu...
Page 312: ...5 Menu 5 1 Menu structure 298 5 1 9 3 Span X Scale Span Span 4 000000 GHz Fig 5 9 3 Span Menu...
Page 323: ...5 Menu 5 1 Menu structure 309 5 1 9 13 Maker Mkr Fig 5 9 13 Marker Menu...
Page 328: ...5 Menu 5 1 Menu structure 314 5 1 10 3 Span X Scale Span Fig 5 10 3 Span Menu...
Page 329: ...5 Menu 5 1 Menu structure 315 5 1 10 4 Bandwidth BW Fig 5 10 4 Bandwidth Menu...
Page 333: ...5 Menu 5 1 Menu structure 319 5 1 10 7 Trace Trace Fig 5 10 7 Trace Menu...
Page 340: ...5 Menu 5 1 Menu structure 326 5 1 10 13 Maker Mkr Fig 5 10 13 Marker Menu...
Page 345: ...5 Menu 5 1 Menu structure 331 5 1 11 3 Span X Scale Span Fig 5 11 3 Span Menu...
Page 357: ...5 Menu 5 1 Menu structure 343 5 1 11 13 Maker Mkr Fig 5 11 13 Marker Menu...
Page 359: ...5 Menu 5 1 Menu structure 345 5 1 11 15 Peak Search Peak Search Fig 5 11 15 Peak Search Menu...
Page 656: ......