8. Specifications and test methods
8.5 Performance characteristics test
668
29)
Set the power meter frequency as 40.05GHz.
30)
Record the 4051 marker reading L
2
and the power meter reading L
1
in the 4051 Auxiliary Form.
31)
Execute Frequency Step Up for the synthesized signal generator and the 4051 analyzer.
32)
Repeat the power meter frequency as the frequency value to be tested next in B8 of the 4051 Auxiliary
Form and repeat step 30 to 31 for each frequency point in the 4051 Auxiliary Form.
Frequency response: (50GHz
~
67GHz)
33)
Set the output frequency of the synthesized signal generator as 50.05GHz and the center frequency of the
4051 analyzer as 50.05GHz.
34)
Set the power meter frequency as 50.05GHz.
35)
Record the 4051 marker reading L
2
and the power meter reading L
1
in the 4051 Auxiliary Form.
36)
Execute Frequency Step Up for the synthesized signal generator and the 4051 analyzer.
37)
Repeat the power meter frequency as the frequency value to be tested next in B8 of the 4051 Auxiliary
Form and repeat step 35 to 36for each frequency point in the 4051 Auxiliary Form.
Frequency response: (3Hz
~
50MHz)
38)
Set the 4051 analyzer as follows:
Center frequency as 50MHz
Span: 100Hz
Resolution bandwidth10 Hz
Marker All Markers Off
39)
Connect the test instrument as shown in Fig. 8.10 and then connect two output interfaces of the power
divider to 50 Ω load and the signal/spectrum analyzer respectively.
Fig. 8.10 Schematic diagram of frequency response test (3Hz ~ 50MHz)
40)
Set Peak Search for the 4051 analyzer and adjust the amplitude of the synthesized signal generator until
the 4051 amplitude reading is closest to the marker reading recorded in step 5.
41)
Set Marker Delta for the 4051 analyzer.
42)
Set the center frequency of the 4051 analyzer and the frequency of the synthesized signal generator as the
frequencies listed in the form and record the Marker Delta reading at each frequency point in the ―relative
50 MHz amplitude error‖ column in B8 of the 4051 Auxiliary Form.
43)
Repeat step 37 for each frequency point in the form.
44)
Record the 50 MHz amplitude error of the first point in ―amplitude error (band 0, 50 MHz ~ 4 GHz)‖ in
B8 of the 4051 Auxiliary Form in ―50 MHz amplitude error‖ under ―amplitude error (band 0, 0 MHz)‖.
Frequency response (pre-amplifier on)
Summary of Contents for 4051 Series
Page 5: ......
Page 6: ......
Page 7: ......
Page 14: ......
Page 18: ......
Page 68: ......
Page 225: ...5 Menu 5 1 Menu structure 211 5 1 3 13 Maker Mkr Fig 5 3 13 Marker Menu...
Page 241: ...5 Menu 5 1 Menu structure 227 5 1 4 13 Maker Mkr Fig 5 4 13 Marker Menu...
Page 259: ...5 Menu 5 1 Menu structure 245 5 1 5 13 Maker Mkr Fig 5 5 13 Marker Menu...
Page 266: ...5 Menu 5 1 Menu structure 252 5 1 6 5 Sweep Sweep Fig 5 6 5 Sweep Menu...
Page 275: ...5 Menu 5 1 Menu structure 261 5 1 6 13 Maker Mkr Fig 5 6 13 Marker Menu...
Page 277: ...5 Menu 5 1 Menu structure 263 5 1 6 15 Peak Search Peak Search Fig 5 6 15 Peak Search Menu...
Page 280: ...5 Menu 5 1 Menu structure 266 5 1 7 3 Span X Scale Span Fig 5 7 3 Span Menu...
Page 291: ...5 Menu 5 1 Menu structure 277 5 1 7 13 Maker Mkr Fig 5 7 13 Marker Menu...
Page 296: ...5 Menu 5 1 Menu structure 282 5 1 8 3 Span X Scale Span Fig 5 8 3 Span Menu...
Page 306: ...5 Menu 5 1 Menu structure 292 5 1 8 12 Marker Marker Fig 5 8 12 Marker Menu...
Page 307: ...5 Menu 5 1 Menu structure 293 5 1 8 13 Maker Mkr Fig 5 8 13 Marker Menu...
Page 309: ...5 Menu 5 1 Menu structure 295 5 1 8 15 Peak Search Peak Search Fig 5 8 15 Peak Search Menu...
Page 312: ...5 Menu 5 1 Menu structure 298 5 1 9 3 Span X Scale Span Span 4 000000 GHz Fig 5 9 3 Span Menu...
Page 323: ...5 Menu 5 1 Menu structure 309 5 1 9 13 Maker Mkr Fig 5 9 13 Marker Menu...
Page 328: ...5 Menu 5 1 Menu structure 314 5 1 10 3 Span X Scale Span Fig 5 10 3 Span Menu...
Page 329: ...5 Menu 5 1 Menu structure 315 5 1 10 4 Bandwidth BW Fig 5 10 4 Bandwidth Menu...
Page 333: ...5 Menu 5 1 Menu structure 319 5 1 10 7 Trace Trace Fig 5 10 7 Trace Menu...
Page 340: ...5 Menu 5 1 Menu structure 326 5 1 10 13 Maker Mkr Fig 5 10 13 Marker Menu...
Page 345: ...5 Menu 5 1 Menu structure 331 5 1 11 3 Span X Scale Span Fig 5 11 3 Span Menu...
Page 357: ...5 Menu 5 1 Menu structure 343 5 1 11 13 Maker Mkr Fig 5 11 13 Marker Menu...
Page 359: ...5 Menu 5 1 Menu structure 345 5 1 11 15 Peak Search Peak Search Fig 5 11 15 Peak Search Menu...
Page 656: ......