8. Specifications and test methods
8.5 Performance characteristics test
663
time. The video bandwidth is guaranteed during design and production by the manufacturer of the analyzer
and can be checked via the video bandwidth setting.
b)
Test procedure
1)
Reset the 4051 analyzer, enter the bandwidth menu and set the video bandwidth from 1 Hz to 3 MHz or as
4, 5, 6, 8 and 10 MHz.
2)
Determine measurement values according to the requirements in Section 8.3.1.12. If all results fall within
the required range, ―compliant‖ should be noted in the 4051 Record Sheet, otherwise it should be
considered as noncompliant.
8.5.1.1.12
11. Other indicators
a)
Test items
Other indicators mainly include the local measurement update rate, center frequency tuning and
conversion rate, number of traces, to represent the fast measurement characteristics of 4051 signal analyzer.
b)
Test block diagram and test apparatus and equipment
Synthesized signal generator ......................................... 1463/E8254A is recommended
Oscilloscope ................................................................................ 4445 is recommended
Real-time spectrum analyzer ................................................... FSVR7 is recommended
Fig .8.7 Testing device of other indicators
c)
Test procedure
1
)
Local measurement update rate: connect apparatus as shown in Fig. 8.7. Set 4051 as spectrum
analysis mode, analysis bandwidth as 3MHz, frequency width as 0Hz and scan time as 1us, use
the oscilloscope to measure the pulse rising time of signal analyzer rear panel trigger output
end, and measure the mean period as the local measurement update rate.
2
)
Center frequency tuning and conversion rate: the center frequency tuning and conversion rate
is the interval time between two 4051 scan processes. Set 4051 as spectrum analysis mode,
scan width as 0Hz, scan time as 1ms and frequency step as 5MHz to respectively set the center
frequency as 1GHz, 5GHz and 27GHz. For the step center frequency, adopt the real-time
spectrum analyzer FSVR to capture the 4051 first local oscillation jump process, and read out
the frequency jump interval time as center frequency tuning and conversion rate.
3
)
Number of traces: press the key "Trace" and soft menu "Trace selection" to select 6 traces.
d)
Test record and data processing
4
)
Record the measuring result under the relevant test items in
《
Record Form
》
.
合成信号发生器
信号
/
频谱分析仪
转接器
转接器
2.4mm
电缆
Summary of Contents for 4051 Series
Page 5: ......
Page 6: ......
Page 7: ......
Page 14: ......
Page 18: ......
Page 68: ......
Page 225: ...5 Menu 5 1 Menu structure 211 5 1 3 13 Maker Mkr Fig 5 3 13 Marker Menu...
Page 241: ...5 Menu 5 1 Menu structure 227 5 1 4 13 Maker Mkr Fig 5 4 13 Marker Menu...
Page 259: ...5 Menu 5 1 Menu structure 245 5 1 5 13 Maker Mkr Fig 5 5 13 Marker Menu...
Page 266: ...5 Menu 5 1 Menu structure 252 5 1 6 5 Sweep Sweep Fig 5 6 5 Sweep Menu...
Page 275: ...5 Menu 5 1 Menu structure 261 5 1 6 13 Maker Mkr Fig 5 6 13 Marker Menu...
Page 277: ...5 Menu 5 1 Menu structure 263 5 1 6 15 Peak Search Peak Search Fig 5 6 15 Peak Search Menu...
Page 280: ...5 Menu 5 1 Menu structure 266 5 1 7 3 Span X Scale Span Fig 5 7 3 Span Menu...
Page 291: ...5 Menu 5 1 Menu structure 277 5 1 7 13 Maker Mkr Fig 5 7 13 Marker Menu...
Page 296: ...5 Menu 5 1 Menu structure 282 5 1 8 3 Span X Scale Span Fig 5 8 3 Span Menu...
Page 306: ...5 Menu 5 1 Menu structure 292 5 1 8 12 Marker Marker Fig 5 8 12 Marker Menu...
Page 307: ...5 Menu 5 1 Menu structure 293 5 1 8 13 Maker Mkr Fig 5 8 13 Marker Menu...
Page 309: ...5 Menu 5 1 Menu structure 295 5 1 8 15 Peak Search Peak Search Fig 5 8 15 Peak Search Menu...
Page 312: ...5 Menu 5 1 Menu structure 298 5 1 9 3 Span X Scale Span Span 4 000000 GHz Fig 5 9 3 Span Menu...
Page 323: ...5 Menu 5 1 Menu structure 309 5 1 9 13 Maker Mkr Fig 5 9 13 Marker Menu...
Page 328: ...5 Menu 5 1 Menu structure 314 5 1 10 3 Span X Scale Span Fig 5 10 3 Span Menu...
Page 329: ...5 Menu 5 1 Menu structure 315 5 1 10 4 Bandwidth BW Fig 5 10 4 Bandwidth Menu...
Page 333: ...5 Menu 5 1 Menu structure 319 5 1 10 7 Trace Trace Fig 5 10 7 Trace Menu...
Page 340: ...5 Menu 5 1 Menu structure 326 5 1 10 13 Maker Mkr Fig 5 10 13 Marker Menu...
Page 345: ...5 Menu 5 1 Menu structure 331 5 1 11 3 Span X Scale Span Fig 5 11 3 Span Menu...
Page 357: ...5 Menu 5 1 Menu structure 343 5 1 11 13 Maker Mkr Fig 5 11 13 Marker Menu...
Page 359: ...5 Menu 5 1 Menu structure 345 5 1 11 15 Peak Search Peak Search Fig 5 11 15 Peak Search Menu...
Page 656: ......