5 Menu
5.2. Description of menu
359
[:sense]:power[:RF]:attenuation?
[:sense]:power[:RF]:attenuation:AUTO OFF|ON|0|1
[:sense]:power[:RF]:attenuation:AUTO?
●
Examples:
:POW:ATT 30 dB
:POW:ATT:AUTO ON
●
Other instructions:
The mechanical attenuator, in general, is set as Auto ON mode. When the reference level changes, the
attenuation will automatically change, but this does not change the reference level. This change aims to ensure
that the maximum signal amplitude of the input mixer is less than or equals to -10dBm. For example, if the
reference level is +23dBm and the mixer input level is -17dBm, the mixer is 40dB (23- 40= -17). The final
purpose is to prevent signal compression. When the mechanical atten is under Auto Off mode, press the
[Mech Atten Auto Man], couple the attenuation to the reflevel again.
Note: the maximum input signal amplitude of the input attenuator (at least 10dB input attenuation) is +30dBm,
otherwise, this power signal will damage the mechanical attenuator or the input mixer.
[Mech Atten Step 2dB 10dB]
●
Menu description:
to control the step of the mechanical attenuator. The step size can only be increased
or decreased by 10dB or 2dB. For 2dB step, the input mixer level can be set based on input signal
optimization to improve measurement dynamic scope.
●
Key path:
【
Ampld/YScale
】
, [Attenuation
▶
], [Mech Atten Step 2 dB 10 dB]on the front panel.
●
Status saved:
saved in the instrument status.
●
Factory preset:
2 dB.
●
Remote Command:
Frequency step auto on/off
[:sense]:power[:RF]:attenuation?
[:sense]:power[:RF]:attenuation:AUTO OFF|ON|0|1
[:sense]:power[:RF]:attenuation:AUTO?
[Max Mixer Level]
●
Menu description:
to set the input maximum mixer level. This level is used as the maximum level limit
of the input mixer for assignment of reference level. The level is equivalent to reference level value
minus the attenuator decrement When the reference level changes, the mechanical attenuator also
changes to ensure that the signal display on the reference level will not be larger than the maximum
mixer level. If the maximum mixer level decreases, the instrument will automatically optimize the
distortion performance under default measurement mode. If the maximum mixer level increases, the
mixer may be damaged by excessive input signal. This parameter cannot be changed unless the user has a
good knowledge of the signal analyzer.
●
Key path:
【
Ampld/YScale
】
, [Attenuation
▶
], [Max Mixer Level] on the front panel.
●
Status saved:
saved in the instrument status.
●
Factory preset:
-10dBm.
●
Setting range:
Min.: -50dBm, Max.: -10dBm.
Summary of Contents for 4051 Series
Page 5: ......
Page 6: ......
Page 7: ......
Page 14: ......
Page 18: ......
Page 68: ......
Page 225: ...5 Menu 5 1 Menu structure 211 5 1 3 13 Maker Mkr Fig 5 3 13 Marker Menu...
Page 241: ...5 Menu 5 1 Menu structure 227 5 1 4 13 Maker Mkr Fig 5 4 13 Marker Menu...
Page 259: ...5 Menu 5 1 Menu structure 245 5 1 5 13 Maker Mkr Fig 5 5 13 Marker Menu...
Page 266: ...5 Menu 5 1 Menu structure 252 5 1 6 5 Sweep Sweep Fig 5 6 5 Sweep Menu...
Page 275: ...5 Menu 5 1 Menu structure 261 5 1 6 13 Maker Mkr Fig 5 6 13 Marker Menu...
Page 277: ...5 Menu 5 1 Menu structure 263 5 1 6 15 Peak Search Peak Search Fig 5 6 15 Peak Search Menu...
Page 280: ...5 Menu 5 1 Menu structure 266 5 1 7 3 Span X Scale Span Fig 5 7 3 Span Menu...
Page 291: ...5 Menu 5 1 Menu structure 277 5 1 7 13 Maker Mkr Fig 5 7 13 Marker Menu...
Page 296: ...5 Menu 5 1 Menu structure 282 5 1 8 3 Span X Scale Span Fig 5 8 3 Span Menu...
Page 306: ...5 Menu 5 1 Menu structure 292 5 1 8 12 Marker Marker Fig 5 8 12 Marker Menu...
Page 307: ...5 Menu 5 1 Menu structure 293 5 1 8 13 Maker Mkr Fig 5 8 13 Marker Menu...
Page 309: ...5 Menu 5 1 Menu structure 295 5 1 8 15 Peak Search Peak Search Fig 5 8 15 Peak Search Menu...
Page 312: ...5 Menu 5 1 Menu structure 298 5 1 9 3 Span X Scale Span Span 4 000000 GHz Fig 5 9 3 Span Menu...
Page 323: ...5 Menu 5 1 Menu structure 309 5 1 9 13 Maker Mkr Fig 5 9 13 Marker Menu...
Page 328: ...5 Menu 5 1 Menu structure 314 5 1 10 3 Span X Scale Span Fig 5 10 3 Span Menu...
Page 329: ...5 Menu 5 1 Menu structure 315 5 1 10 4 Bandwidth BW Fig 5 10 4 Bandwidth Menu...
Page 333: ...5 Menu 5 1 Menu structure 319 5 1 10 7 Trace Trace Fig 5 10 7 Trace Menu...
Page 340: ...5 Menu 5 1 Menu structure 326 5 1 10 13 Maker Mkr Fig 5 10 13 Marker Menu...
Page 345: ...5 Menu 5 1 Menu structure 331 5 1 11 3 Span X Scale Span Fig 5 11 3 Span Menu...
Page 357: ...5 Menu 5 1 Menu structure 343 5 1 11 13 Maker Mkr Fig 5 11 13 Marker Menu...
Page 359: ...5 Menu 5 1 Menu structure 345 5 1 11 15 Peak Search Peak Search Fig 5 11 15 Peak Search Menu...
Page 656: ......