User Manual
74
i Series / iX Series
4.2.12 OUTPUT IMPEDANCE Menu
[
iX Series only
]
Figure 4-25: OUTPUT IMPEDANCE menu
The iX Series offers programmable output impedance on selected models. This allows the user
to simulate line impedance conditions by programming resistive and inductive elements of the
AC source‟s output impedance.
Models on which this function is available are:
3001iX, 5001iX, 9003iX, 15003iX.
Note: On 9003iX-MODE and 15003iX-MODE systems, programmable impedance function
is only available in 3 phase mode of operation.
Alternative, one of the available OMNI lumped reference impedance options may be used to
increase the AC source output impedance for applications such as IEC 61000-3-3 Flicker testing.
The OUTPUT IMPEDANCE menu can be accessed from MENU 3. The following fields are
available in this menu:
Entry
Description
RESISTIVE
Sets the resistive component of the output impedance. This
value can be set from 17 m to 1000 m .
(The lower limits for the resistive and inductive components may
vary on different AC source models and are set as part of the
source calibration.)
INDUCTIVE
Sets the inductive component of the output impedance. This
value can be set from 230 H to 1000 H.
(The lower limits for the resistive and inductive components may
vary on different AC source models and are set as part of the
source calibration.)
SET MINIMUM
Selecting this field and pressing the ENTER key sets both
impedance components to their minimum values.
SET IEC1000-3-3
Selecting this field and pressing the ENTER key sets the output
impedance to the IEC 725 reference impedance values required
for IEC 1000-3-3 flicker testing.
Summary of Contents for 10001i
Page 2: ......
Page 3: ......
Page 6: ...ii This page intentionally left blank...
Page 38: ...User Manual 24 i Series iX Series Figure 3 5 Rear Panel View for the 3001i 3001iX...
Page 39: ...User Manual i Series iX Series 25 Figure 3 6 Rear Panel View for the 5001i 5001iX...
Page 43: ...User Manual i Series iX Series 29 Figure 3 8 Functional Test Setup...
Page 44: ...User Manual 30 i Series iX Series Figure 3 9 Single Phase 10000 VA System 10001iX i...
Page 46: ...User Manual 32 i Series iX Series Figure 3 11 Single Phase 15000 VA System 15001iX i...
Page 48: ...User Manual 34 i Series iX Series Figure 3 13 Connection With MODE Option...
Page 118: ...User Manual 104 i Series iX Series Figure 5 2 Power Source Module Block Diagram...
Page 121: ...User Manual i Series iX Series 107 Figure 5 3 5001i Internal Layout...
Page 122: ...User Manual 108 i Series iX Series Figure 5 4 Logic Board LED s...
Page 124: ...User Manual 110 i Series iX Series Figure 5 5 AC Power Stage Layout...
Page 125: ...User Manual i Series iX Series 111 Figure 5 6 AC Control Logic Block Diagram...
Page 138: ...User Manual 124 i Series iX Series Figure 6 3 Adjustment Location...
Page 152: ...User Manual 138 i Series iX Series Figure 9 4 Voltage Modulation...
Page 219: ...User Manual i Series iX Series 205 Figure 9 36 Example Connection With 5001iX and EOS 1...
Page 221: ...User Manual i Series iX Series 207 Figure 9 38 15003iX CTS EOS3 LR3...
Page 222: ...User Manual 208 i Series iX Series Figure 9 39 15003iX 3 EOS3...
Page 233: ...User Manual i Series iX Series 219 Figure 9 41 Example Connection With MODE iX...
Page 240: ...User Manual 226 i Series iX Series Figure 9 42 Example Connections With OMNI 1 18i...
Page 241: ...User Manual i Series iX Series 227 Figure 9 43 Example Connections With OMNI 3 18i...