6.5 Routine Measurement Calibration
Connect the test equipment to the power source as shown in Figure 6-2. If the power system is
a multi-phase system with one controller, the DVM for calibrating the measurement voltage
should always be connected to the Remote Sense connector (TB3) on the Phase A power
source.
The shunt must be connected to the power source as shown in Figure 6-2. If the Current
measurement can‟t be successfully performed, adjust the Current Measurement Pot. This
adjustment is described in the Non-routine Calibration section of this manual. If the DC current
measurement displays more than 70 counts on the display, perform the non-routine current
monitor adjustment.
Connect the load to the output. Use the 10 milliohm current shunt in series with the load to
measure the AC and DC load current. When programming a DC load always program the output
voltage to 0 volts before changing the output load. This will prevent load switch contacts from
being damaged.
To calibrate all measurement functions, the desired value for the measurement value of current
or voltage must be entered for the corresponding calibration value. Make the indicated
adjustments by typing in the desired display value. This should be the value indicated by the
external DVM. If a 10 milliohm current shunt is used for current, 300 millivolts represents 30
amps.
The Measurement Calibration Table is a summary of the measurement calibration procedure.
The following text is a detailed explanation of the procedure.
PARAMETER
POWER SYSTEM
3001
5001
10001
15001
15003
AC Current Full-Scale
6.8 , 3KW
4.1 , 5KW
2.1 , 10KW
1.35 , 15KW
DC Current Full-Scale
13.5 , 1.5KW
8 , 2.5KW
4 , 5KW
2.7 , 7.5KW
Table 6-2: Calibration Load For Each Phase
Summary of Contents for 10001i
Page 2: ......
Page 3: ......
Page 6: ...ii This page intentionally left blank...
Page 38: ...User Manual 24 i Series iX Series Figure 3 5 Rear Panel View for the 3001i 3001iX...
Page 39: ...User Manual i Series iX Series 25 Figure 3 6 Rear Panel View for the 5001i 5001iX...
Page 43: ...User Manual i Series iX Series 29 Figure 3 8 Functional Test Setup...
Page 44: ...User Manual 30 i Series iX Series Figure 3 9 Single Phase 10000 VA System 10001iX i...
Page 46: ...User Manual 32 i Series iX Series Figure 3 11 Single Phase 15000 VA System 15001iX i...
Page 48: ...User Manual 34 i Series iX Series Figure 3 13 Connection With MODE Option...
Page 118: ...User Manual 104 i Series iX Series Figure 5 2 Power Source Module Block Diagram...
Page 121: ...User Manual i Series iX Series 107 Figure 5 3 5001i Internal Layout...
Page 122: ...User Manual 108 i Series iX Series Figure 5 4 Logic Board LED s...
Page 124: ...User Manual 110 i Series iX Series Figure 5 5 AC Power Stage Layout...
Page 125: ...User Manual i Series iX Series 111 Figure 5 6 AC Control Logic Block Diagram...
Page 138: ...User Manual 124 i Series iX Series Figure 6 3 Adjustment Location...
Page 152: ...User Manual 138 i Series iX Series Figure 9 4 Voltage Modulation...
Page 219: ...User Manual i Series iX Series 205 Figure 9 36 Example Connection With 5001iX and EOS 1...
Page 221: ...User Manual i Series iX Series 207 Figure 9 38 15003iX CTS EOS3 LR3...
Page 222: ...User Manual 208 i Series iX Series Figure 9 39 15003iX 3 EOS3...
Page 233: ...User Manual i Series iX Series 219 Figure 9 41 Example Connection With MODE iX...
Page 240: ...User Manual 226 i Series iX Series Figure 9 42 Example Connections With OMNI 1 18i...
Page 241: ...User Manual i Series iX Series 227 Figure 9 43 Example Connections With OMNI 3 18i...