User Manual
i Series / iX Series
323
Number
Message String
Cause
Remedy
6
"Limit memory lost"
Hardware
configuration settings
lost.
Contact CI service department at
to obtain
instructions on restoring configuration
data.
7
"System memory lost"
Memory corrupted.
Recycle power.
8
"Calibration memory
lost"
Calibration data lost.
Contact CI service department at
to obtain
instructions on restoring calibration
data or recalibrate unit.
9
"Start angle must be
first sequence"
Start phase angle in
wrong place..
Start phase angles can only
programmed at the start of a transient
list. Once a transient is in progress,
phase angle cannot be changed.
10
"Illegal for DC"
Operation not
possible in DC mode.
Switch to AC or AC+DC mode.
11
"Duplicate sequence"
N/A
N/A
12
"Too many sequence"
N/A
N/A
13
"Missing list
parameter"
One or more
transient list
parameters missing.
Check programmed lists.
14
"Voltage peak error "
Peak voltage
exceeds internal bus
voltage
This error may occur when selecting
user defined wave shapes with higher
crest factors. Reduce programmed
RMS value.
15
"Slew time exceed
dwell"
Time needed to slew
to final value is less
than dwell time.
Check dwell times in transient list
settings. Increase dwell time or
change slew rate for affected
parameter.
16
"Illegal during
transient"
Operation requested
not available while
transient is running.
Wait till transient execution is
completed or abort transient
execution first.
17
"Output relay must be
closed"
Operation not
possible with open
relay
Close relay before attempting
operation. E.g. transient execution
requires output relay to be closed.
18
"Trans. duration less
then 1msec"
Dwell time below
minimum or 1 msec
Increase dwell time to at least 1
msec.
19
"Clock and sync must
be internal"
Operation not
possible with external
clock
Switch to internal sync. (Default)
20
"Input buffer full"
Too much data
received.
Break up data in smaller blocks.
21
"EOS Fault"
Hardware error
reported by EOS
option
Cycle power on EOS to reset error. If
error persist, contact Contact CI
service department at
for repair.
22
"Waveform harmonics
limit"
Harmonic contents of
user defined wave
shape is too high and
could damage
amplifier output
stage.
Reduce harmonic content or reduce
fundamental frequency programmed.
Summary of Contents for 10001i
Page 2: ......
Page 3: ......
Page 6: ...ii This page intentionally left blank...
Page 38: ...User Manual 24 i Series iX Series Figure 3 5 Rear Panel View for the 3001i 3001iX...
Page 39: ...User Manual i Series iX Series 25 Figure 3 6 Rear Panel View for the 5001i 5001iX...
Page 43: ...User Manual i Series iX Series 29 Figure 3 8 Functional Test Setup...
Page 44: ...User Manual 30 i Series iX Series Figure 3 9 Single Phase 10000 VA System 10001iX i...
Page 46: ...User Manual 32 i Series iX Series Figure 3 11 Single Phase 15000 VA System 15001iX i...
Page 48: ...User Manual 34 i Series iX Series Figure 3 13 Connection With MODE Option...
Page 118: ...User Manual 104 i Series iX Series Figure 5 2 Power Source Module Block Diagram...
Page 121: ...User Manual i Series iX Series 107 Figure 5 3 5001i Internal Layout...
Page 122: ...User Manual 108 i Series iX Series Figure 5 4 Logic Board LED s...
Page 124: ...User Manual 110 i Series iX Series Figure 5 5 AC Power Stage Layout...
Page 125: ...User Manual i Series iX Series 111 Figure 5 6 AC Control Logic Block Diagram...
Page 138: ...User Manual 124 i Series iX Series Figure 6 3 Adjustment Location...
Page 152: ...User Manual 138 i Series iX Series Figure 9 4 Voltage Modulation...
Page 219: ...User Manual i Series iX Series 205 Figure 9 36 Example Connection With 5001iX and EOS 1...
Page 221: ...User Manual i Series iX Series 207 Figure 9 38 15003iX CTS EOS3 LR3...
Page 222: ...User Manual 208 i Series iX Series Figure 9 39 15003iX 3 EOS3...
Page 233: ...User Manual i Series iX Series 219 Figure 9 41 Example Connection With MODE iX...
Page 240: ...User Manual 226 i Series iX Series Figure 9 42 Example Connections With OMNI 1 18i...
Page 241: ...User Manual i Series iX Series 227 Figure 9 43 Example Connections With OMNI 3 18i...