User Manual
254
i Series / iX Series
9.8.2.3 Specific Test Limitations and Discrepancies
While every effort has been made to provide comprehensive coverage of the Mil-Std 704 tests,
some limitations apply, largely due to hardware limitations of the iX power source. The following
table lists specific implementation issues by test number. The deviation from the standard
requirement is shown in the “Actual setting” column. The reason for the deviation is shown in the
last column.
Rev
Test
Std requirement
Actual setting
Reason
A
SAC104
.5Vp-v modulation (.3536V RMS)
.4V RMS modulation
Resolution
A
SAC104
3.5Vp-v modulation (2.475V RMS)
2.4V RMS modulation
Resolution
A
TAC104
.5Vp-v modulation (.3536V RMS)
.4V RMS modulation
Resolution
A
TAC104
3.5Vp-v modulation (2.475V RMS)
2.4V RMS modulation
Resolution
A-F
SAC106
50kHz
Test skipped
Requires additional equipment
A-F
TAC106
50kHz
Test skipped
Requires additional equipment
A-F
LDC103
50kHz
Test skipped
Requires additional equipment
A-F
LDC104
16.8kHz
Test skipped
Requires additional equipment
B-F
SAC104
.375V RMS modulation
.4V RMS modulation
Resolution
B-F
SAC104
2.5V RMS modulation
2.4V RMS modulation
Resolution
B-F
TAC104
.375V RMS modulation
.4V RMS modulation
Resolution
B-F
TAC104
2.5V RMS modulation
2.4V RMS modulation
Resolution
B-F
HDC103
50kHz
Test skipped
Requires additional equipment
B-F
HDC104
16.8kHz
Test skipped
Requires additional equipment
F
SVF104
.375V RMS modulation
.4V RMS modulation
Resolution
F
SVF104
2.5V RMS modulation
2.4V RMS modulation
Resolution
F
SVF106
50kHz
Test skipped
Requires additional equipment
F
TVF104
.375V RMS modulation
.4V RMS modulation
Resolution
F
TVF104
2.5V RMS modulation
2.4V RMS modulation
Resolution
F
TVF106
50kHz
Test skipped
Requires additional equipment
F
SXF104
.375V RMS modulation
.4V RMS modulation
Resolution
F
SXF104
2.5V RMS modulation
2.4V RMS modulation
Resolution
F
SXF106
50kHz
Test skipped
Requires additional equipment
F
SXF109
Dwell time = 1 cycle (16.666667 msec) Dwell time = 17msec
Resolution
F
SXF201
Dwell time = 1 cycle (16.666667 msec) Dwell time = 17msec
Resolution
F
SXF302
Dwell time = 1 cycle (16.666667 msec) Dwell time = 17msec
Resolution
F
HDC105
1
330VDC
300VDC limit (test written but will not run) Hardware limit DC range i/iX
F
HDC302
3
350VDC
300VDC limit (test written but will not run) Hardware limit DC range i/iX
B, C HDC105
475VDC
300VDC limit (test written but will not run) Hardware limit DC range i/iX
B, C HDC302
475VDC
300VDC limit (test written but will not run) Hardware limit DC range i/iX
B, C SAC105
+/-1.75Hz modulation
+/-2Hz modulation
Resolution
B, C SAC105
+/-0.85Hz modulation
+/-0.9Hz modulation
Resolution
B, C SAC105
+/-0.58Hz modulation
+/-0.6Hz modulation
Resolution
B, C TAC105
+/-1.75Hz modulation
+/-2Hz modulation
Resolution
B, C TAC105
+/-0.85Hz modulation
+/-0.9Hz modulation
Resolution
B, C TAC105
+/-0.58Hz modulation
+/-0.6Hz modulation
Resolution
1
Note:
To perform MIL-STD-704F tests HDC105-1, -3 and HDC302-1, -3, it is advised to use the firmware revision
E tests TRANSIENT and ABNORMAL respectively. The
–704 firmware implementation overrides the DC voltage
range hardware limit making it possible for these tests to be performed. See section 9.8.1.8. The same does not
apply to the equivalent rev B or rev C tests, which require 475VDC. The 475 VDC level is not supported on the i/iX
Series.
Summary of Contents for 10001i
Page 2: ......
Page 3: ......
Page 6: ...ii This page intentionally left blank...
Page 38: ...User Manual 24 i Series iX Series Figure 3 5 Rear Panel View for the 3001i 3001iX...
Page 39: ...User Manual i Series iX Series 25 Figure 3 6 Rear Panel View for the 5001i 5001iX...
Page 43: ...User Manual i Series iX Series 29 Figure 3 8 Functional Test Setup...
Page 44: ...User Manual 30 i Series iX Series Figure 3 9 Single Phase 10000 VA System 10001iX i...
Page 46: ...User Manual 32 i Series iX Series Figure 3 11 Single Phase 15000 VA System 15001iX i...
Page 48: ...User Manual 34 i Series iX Series Figure 3 13 Connection With MODE Option...
Page 118: ...User Manual 104 i Series iX Series Figure 5 2 Power Source Module Block Diagram...
Page 121: ...User Manual i Series iX Series 107 Figure 5 3 5001i Internal Layout...
Page 122: ...User Manual 108 i Series iX Series Figure 5 4 Logic Board LED s...
Page 124: ...User Manual 110 i Series iX Series Figure 5 5 AC Power Stage Layout...
Page 125: ...User Manual i Series iX Series 111 Figure 5 6 AC Control Logic Block Diagram...
Page 138: ...User Manual 124 i Series iX Series Figure 6 3 Adjustment Location...
Page 152: ...User Manual 138 i Series iX Series Figure 9 4 Voltage Modulation...
Page 219: ...User Manual i Series iX Series 205 Figure 9 36 Example Connection With 5001iX and EOS 1...
Page 221: ...User Manual i Series iX Series 207 Figure 9 38 15003iX CTS EOS3 LR3...
Page 222: ...User Manual 208 i Series iX Series Figure 9 39 15003iX 3 EOS3...
Page 233: ...User Manual i Series iX Series 219 Figure 9 41 Example Connection With MODE iX...
Page 240: ...User Manual 226 i Series iX Series Figure 9 42 Example Connections With OMNI 1 18i...
Page 241: ...User Manual i Series iX Series 227 Figure 9 43 Example Connections With OMNI 3 18i...