User Manual
180
i Series / iX Series
9.2 IEC 61000-4-11 Option
9.2.1 General
The IEC1000-4-11 option is capable of performing IEC1000-4 section 11 voltage dips, short
interruptions and voltage variations immunity tests. On three-phase iX/i Configurations, the user
can select one, two or all three phases to be active during the IEC1000-4-11 tests in this
configuration.
9.2.2 Standard Revisions and EUT Classes
The
–411 option supports both the first (1994-06) and the second edition (2004-03) of the IEC
61000-4-11 test standard as of firmware revision 2.38. Older firmware revisions only support the
first edition. Contact
for upgrade information. If Edition 2.0 is supported, the
standard revision can be selected when using the Cigui32 Windows program. From the front
panel, there is no need to select the revision but data values can be set that support either
version of the standard.
Generic tests files are distributed with the CIGui32 program for both editions of the test standard.
Files applicable to Edition 2.0 have ED20 in their file name. Do not mix these files, as the data
setup will not be correct if you do. To load a test file, select the Mode (Dips or Vars) and test
standard revision first, then use the File, Open menu to load the test parameters. Test
parameters can be a function of the EUT class. The different files provided with the program
cover the various EUT classes. The relevant EUT class 1, 2, 3 or X is listed in the file names.
When using front panel operation, the user has to set levels for each individual test step in
SINGLE mode.
9.2.3 Initial Setup
The user must set the operating frequency and voltage and close the output relay prior to the
start of test. It is possible to change the normal voltage (Ut) from the IEC1000-4-11 menus
before running each test. To run any IEC 61000-4-11 test, the IEC411 state must be turned on.
If an EOS1 or EOS3 is detected, it will be put in ACTIVE mode as long as the IEC411 state is
ON.
9.2.4 Tests Performed
DIPS AND INTERRUPTIONS
1. Run All
Run predefined sequence of tests.
2. Run Single
Run user defined test.
VOLTAGE VARIATIONS
1. Run All
Run predefined sequence of tests.
2. Run Single
Run user defined test.
Summary of Contents for 10001i
Page 2: ......
Page 3: ......
Page 6: ...ii This page intentionally left blank...
Page 38: ...User Manual 24 i Series iX Series Figure 3 5 Rear Panel View for the 3001i 3001iX...
Page 39: ...User Manual i Series iX Series 25 Figure 3 6 Rear Panel View for the 5001i 5001iX...
Page 43: ...User Manual i Series iX Series 29 Figure 3 8 Functional Test Setup...
Page 44: ...User Manual 30 i Series iX Series Figure 3 9 Single Phase 10000 VA System 10001iX i...
Page 46: ...User Manual 32 i Series iX Series Figure 3 11 Single Phase 15000 VA System 15001iX i...
Page 48: ...User Manual 34 i Series iX Series Figure 3 13 Connection With MODE Option...
Page 118: ...User Manual 104 i Series iX Series Figure 5 2 Power Source Module Block Diagram...
Page 121: ...User Manual i Series iX Series 107 Figure 5 3 5001i Internal Layout...
Page 122: ...User Manual 108 i Series iX Series Figure 5 4 Logic Board LED s...
Page 124: ...User Manual 110 i Series iX Series Figure 5 5 AC Power Stage Layout...
Page 125: ...User Manual i Series iX Series 111 Figure 5 6 AC Control Logic Block Diagram...
Page 138: ...User Manual 124 i Series iX Series Figure 6 3 Adjustment Location...
Page 152: ...User Manual 138 i Series iX Series Figure 9 4 Voltage Modulation...
Page 219: ...User Manual i Series iX Series 205 Figure 9 36 Example Connection With 5001iX and EOS 1...
Page 221: ...User Manual i Series iX Series 207 Figure 9 38 15003iX CTS EOS3 LR3...
Page 222: ...User Manual 208 i Series iX Series Figure 9 39 15003iX 3 EOS3...
Page 233: ...User Manual i Series iX Series 219 Figure 9 41 Example Connection With MODE iX...
Page 240: ...User Manual 226 i Series iX Series Figure 9 42 Example Connections With OMNI 1 18i...
Page 241: ...User Manual i Series iX Series 227 Figure 9 43 Example Connections With OMNI 3 18i...