i Series / iX Series
Number
Message String
Cause
Remedy
error
-311
"Memory error"
Waveform memory
checksum error.
May be the result of incomplete user-
defined waveform download. Check
interface and try downloading
waveform again. Successful
download may clear this error
condition.
Alternatively, use TRAC:DEL ALL
command to clear waveform memory.
-314
"Save/recall memory
lost"
User setup register
contents lost
Store setup in same register again.
-315
"Configuration memory
lost"
Hardware
configuration settings
lost.
Contact CI service department at
to obtain
instructions on restoring configuration
data.
-330
"Self-test failed"
Internal error
Contact CI service department at
-350
"Queue overflow"
Message queue full.
Too many message. Read status
using SYST:ERR query until 0, "No
Error" is received indicating queue
empty.
-400
"Query error"
Unable to complete
query.
Check programming manual for
correct query format and parameters
-410
"Query
INTERRUPTED"
Query issued but
response not read.
Check application program for correct
flow. Response must be read after
each query to avoid this error.
-420
"Query
UNTERMINATED"
Query incomplete.
Check for terminator after query
command.
-430
"Query
DEADLOCKED"
Query cannot be
completed
Check application program for
multiple queries
-440
"Query
UNTERMINATED"
Query incomplete.
Check for terminator after query
command.
1
"Output volt fault"
Output voltage does
not match
programmed value.
Load exceeds current limit and unit is
in Constant Voltage (CV) mode of
operation. Reduce load or increase
CL setting
Output voltage is driven above
programmed voltage by external
influence (Load, voltage kickback,
etc.)
2
"Current limit fault"
Current limit
exceeded.
Load exceeds current limit and unit is
in Constant Voltage (CV) mode of
operation. Reduce load or increase
CL setting
3
"Temperature fault"
Temperature of heat
sink too high.
Reduce load. Ensure proper air flow
and exhaust clearance. Check fan(s)
for operation.
4
"External sync. error"
Could not sync to
external sync signal.
External sync signal missing,
disconnected or out of range.
5
"Initial memory lost"
Power on settings
could not be recalled.
Save power on settings again to
overwrite old content.
Summary of Contents for 10001i
Page 2: ......
Page 3: ......
Page 6: ...ii This page intentionally left blank...
Page 38: ...User Manual 24 i Series iX Series Figure 3 5 Rear Panel View for the 3001i 3001iX...
Page 39: ...User Manual i Series iX Series 25 Figure 3 6 Rear Panel View for the 5001i 5001iX...
Page 43: ...User Manual i Series iX Series 29 Figure 3 8 Functional Test Setup...
Page 44: ...User Manual 30 i Series iX Series Figure 3 9 Single Phase 10000 VA System 10001iX i...
Page 46: ...User Manual 32 i Series iX Series Figure 3 11 Single Phase 15000 VA System 15001iX i...
Page 48: ...User Manual 34 i Series iX Series Figure 3 13 Connection With MODE Option...
Page 118: ...User Manual 104 i Series iX Series Figure 5 2 Power Source Module Block Diagram...
Page 121: ...User Manual i Series iX Series 107 Figure 5 3 5001i Internal Layout...
Page 122: ...User Manual 108 i Series iX Series Figure 5 4 Logic Board LED s...
Page 124: ...User Manual 110 i Series iX Series Figure 5 5 AC Power Stage Layout...
Page 125: ...User Manual i Series iX Series 111 Figure 5 6 AC Control Logic Block Diagram...
Page 138: ...User Manual 124 i Series iX Series Figure 6 3 Adjustment Location...
Page 152: ...User Manual 138 i Series iX Series Figure 9 4 Voltage Modulation...
Page 219: ...User Manual i Series iX Series 205 Figure 9 36 Example Connection With 5001iX and EOS 1...
Page 221: ...User Manual i Series iX Series 207 Figure 9 38 15003iX CTS EOS3 LR3...
Page 222: ...User Manual 208 i Series iX Series Figure 9 39 15003iX 3 EOS3...
Page 233: ...User Manual i Series iX Series 219 Figure 9 41 Example Connection With MODE iX...
Page 240: ...User Manual 226 i Series iX Series Figure 9 42 Example Connections With OMNI 1 18i...
Page 241: ...User Manual i Series iX Series 227 Figure 9 43 Example Connections With OMNI 3 18i...