User Manual
298
i Series / iX Series
Control / Display
Description
past is maximum and will no longer update as the total test time
is determined by the operator now and not the program.
Revert to
Steady State
on Abort
When selected, the output of the power source will be set back
to the steady state settings in effect before the start of the test
when the operator aborts a test. Normally, the output of the
source at the end of a test run is determined by the settings in
the test sequence file loaded.
Options
This section contains additional options that can be set by the operator as
needed. The following options are available:
Auto Acquire
EUT Data
after
execution
When set, this option causes the GUI program to initiate an EUT
performance data measurement sequence automatically at the
end of a test (normal mode) or test step (single step mode).
Show Phase
B, C data
For three-
phase EUT‟s, data for phases B and C can be
displayed as well. In most cases, the settings for all three
phases are the same and there is no need to display them all as
it takes up screen space. However, this is not the case for
voltage imbalance tests. This is a display feature only and has
no effect on the actual test. Note however, that if the power
source is in single-phase mode, settings for phase B and C will
be ignored. The phase mode of the power source can only be
changed from the main GUI screen.
Test Status
This area of the Test Control tab displays test execution status for each individual
test step as well as the complete test. The top section shows the complete test
time and what part has been completed so far. The bottom section shows the
same information for the present test step.
Test Description
The area at the top of the Test Control tab is used to display additional
information about the selected test tab. It is for information purposes only and has
no effect on the actual test execution.
Table Header
The table header is shown against a yellow background and shows the table from
which this test sequence was selected.
Test Sequence Data
Grid
The test sequence data grid displays the specific parameters used to program the
power source. This data implements the required tests. The parameters are
loaded for the referenced ini file in the test table on the Test Selection tab. It is
possible to edit this information if needed which allows variations on any of these
tests to be made. Any changes can be saved to a new file. The standard provided
test files are write protected however and should not be overwritten.
The following columns/data fields are available in this table for each test step
(row).
Step number
Sequential number from 1 through end of sequence.
Freq
Test frequency in Hz
FSlew
Frequency slew rate in Hz/sec
Volt A
Test voltage for phase/output A in Vrms or Vdc depending on
mode.
VSlew A
Voltage slew rate in V/sec
Func A
Waveform function. Default is SINE for AC mode and DC for DC
mode. Arbitrary waveform can be used in AC mode if this
capability is supported by the power source.
Volt B
Test voltage for phase/output B in Vrms or Vdc depending on
mode.
VSlew B
Voltage slew rate in V/sec
Func B
Waveform function. Default is SINE for AC mode and DC for DC
mode. Arbitrary waveform can be used in AC mode if this
capability is supported by the power source.
Volt C
Test voltage for phase/output C in Vrms or Vdc depending on
mode.
VSlew C
Voltage slew rate in V/sec
Func C
Waveform function. Default is SINE for AC mode and DC for DC
mode. Arbitrary waveform can be used in AC mode if this
Summary of Contents for 10001i
Page 2: ......
Page 3: ......
Page 6: ...ii This page intentionally left blank...
Page 38: ...User Manual 24 i Series iX Series Figure 3 5 Rear Panel View for the 3001i 3001iX...
Page 39: ...User Manual i Series iX Series 25 Figure 3 6 Rear Panel View for the 5001i 5001iX...
Page 43: ...User Manual i Series iX Series 29 Figure 3 8 Functional Test Setup...
Page 44: ...User Manual 30 i Series iX Series Figure 3 9 Single Phase 10000 VA System 10001iX i...
Page 46: ...User Manual 32 i Series iX Series Figure 3 11 Single Phase 15000 VA System 15001iX i...
Page 48: ...User Manual 34 i Series iX Series Figure 3 13 Connection With MODE Option...
Page 118: ...User Manual 104 i Series iX Series Figure 5 2 Power Source Module Block Diagram...
Page 121: ...User Manual i Series iX Series 107 Figure 5 3 5001i Internal Layout...
Page 122: ...User Manual 108 i Series iX Series Figure 5 4 Logic Board LED s...
Page 124: ...User Manual 110 i Series iX Series Figure 5 5 AC Power Stage Layout...
Page 125: ...User Manual i Series iX Series 111 Figure 5 6 AC Control Logic Block Diagram...
Page 138: ...User Manual 124 i Series iX Series Figure 6 3 Adjustment Location...
Page 152: ...User Manual 138 i Series iX Series Figure 9 4 Voltage Modulation...
Page 219: ...User Manual i Series iX Series 205 Figure 9 36 Example Connection With 5001iX and EOS 1...
Page 221: ...User Manual i Series iX Series 207 Figure 9 38 15003iX CTS EOS3 LR3...
Page 222: ...User Manual 208 i Series iX Series Figure 9 39 15003iX 3 EOS3...
Page 233: ...User Manual i Series iX Series 219 Figure 9 41 Example Connection With MODE iX...
Page 240: ...User Manual 226 i Series iX Series Figure 9 42 Example Connections With OMNI 1 18i...
Page 241: ...User Manual i Series iX Series 227 Figure 9 43 Example Connections With OMNI 3 18i...