User Manual
188
i Series / iX Series
9.3.4 Front Panel Entry
While it is possible to perform IEC 61000-4-13 testing from the front panel of the iX Series AC
power source, it is recommended to use the provided Cigui32 Windows program for report
generation. This also provides a more convenient way to perform Class 1 and User class tests
as test levels can be saved to disk.
To perform a test from the keyboard, select the APPLICATIONS screen from the MENU 2
screen. The APPLICATIONS screen will appear as shown in Figure 9-24.
Figure 9-24: Application menu
Scroll to the IEC 61000-4-13 entry using the up and down cursor keys. Press the ENTER key to
select the IEC 61000-4-13 main menu. The screen will appear as shown in Figure 9-25.
Figure 9-25: IEC 61000-4-13 Menu
9.3.4.1 IEC 61000-4-13 Screen Parameters
The IEC 61000-4-13 screen has the following common fields for all IEC 61000-4-13 test groups.
STATE
The STATE field must be set to ON before the test can run. Initial setup conditions must be met
to run the test.
RUN
The RUN field has the following selections:
1. ALL
Run all IEC 61000-4-13.
2. GROUP
Run the selected group.
3. RANGE
Run the selected range for the selected group.
4. POINT
Run the selected frequency for the selected range and selected group.
Note: The RANGE and POINT do not apply to all groups. See individual groups for details.
When the test is running, this field will have the following selection:
1. ABORT
The test will stop running and all parameters will return to the state prior to the
start of the test.
2. PAUSE
The test will pause at the time the ENTER key is pressed while the PAUSE field
is selected. The field will change to RESUME. When the ENTER key is
pressed again, the test will resume from the point it stopped.
Summary of Contents for 10001i
Page 2: ......
Page 3: ......
Page 6: ...ii This page intentionally left blank...
Page 38: ...User Manual 24 i Series iX Series Figure 3 5 Rear Panel View for the 3001i 3001iX...
Page 39: ...User Manual i Series iX Series 25 Figure 3 6 Rear Panel View for the 5001i 5001iX...
Page 43: ...User Manual i Series iX Series 29 Figure 3 8 Functional Test Setup...
Page 44: ...User Manual 30 i Series iX Series Figure 3 9 Single Phase 10000 VA System 10001iX i...
Page 46: ...User Manual 32 i Series iX Series Figure 3 11 Single Phase 15000 VA System 15001iX i...
Page 48: ...User Manual 34 i Series iX Series Figure 3 13 Connection With MODE Option...
Page 118: ...User Manual 104 i Series iX Series Figure 5 2 Power Source Module Block Diagram...
Page 121: ...User Manual i Series iX Series 107 Figure 5 3 5001i Internal Layout...
Page 122: ...User Manual 108 i Series iX Series Figure 5 4 Logic Board LED s...
Page 124: ...User Manual 110 i Series iX Series Figure 5 5 AC Power Stage Layout...
Page 125: ...User Manual i Series iX Series 111 Figure 5 6 AC Control Logic Block Diagram...
Page 138: ...User Manual 124 i Series iX Series Figure 6 3 Adjustment Location...
Page 152: ...User Manual 138 i Series iX Series Figure 9 4 Voltage Modulation...
Page 219: ...User Manual i Series iX Series 205 Figure 9 36 Example Connection With 5001iX and EOS 1...
Page 221: ...User Manual i Series iX Series 207 Figure 9 38 15003iX CTS EOS3 LR3...
Page 222: ...User Manual 208 i Series iX Series Figure 9 39 15003iX 3 EOS3...
Page 233: ...User Manual i Series iX Series 219 Figure 9 41 Example Connection With MODE iX...
Page 240: ...User Manual 226 i Series iX Series Figure 9 42 Example Connections With OMNI 1 18i...
Page 241: ...User Manual i Series iX Series 227 Figure 9 43 Example Connections With OMNI 3 18i...