båÖäáëÜ
59 38 399 D3352
D3352.076.01.13.02
07.2008
2 – 11
Tab 2
2.5 List of error messages
2.5
E6
06
21
Hardware signal of release button not
detected
z
Check cable L5 (fiber optic cable),
replace if necessary
z
Replace board DX1
z
Replace the X-ray tube assembly
E6
06
22
Broken temperature sensor
z
Replace the X-ray tube assembly
E3
06
23
Hardware signal of release button
applied during power-on
z
Check cable L5:
– Switch unit OFF
– Pull cable L5 off of tube assembly
– Switch the unit ON
– Perform optical check of L5:
- If light is visible: Replace board DX1
- If no light is visible, replace the tube as-
sembly
E5
06
30
Total radiation time exceeded
If a CAN bus error had been reported before...
z
Checking the CAN bus
z
Please contact the Sirona Customer Service
Center (CSC) to find out whether a bugfix by
means of a software update is possible and per-
form such an update if necessary.
E5
06
31
Partial radiation time exceeded
If a CAN bus error had been reported before...
z
Checking the CAN bus
z
Please contact the Sirona Customer Service
Center (CSC) to find out whether a bugfix by
means of a software update is possible and per-
form such an update if necessary.
E5
06
32
Minimum preheating time not
observed
If a CAN bus error had been reported before...
z
Checking the CAN bus
z
Please contact the Sirona Customer Service
Center (CSC) to find out whether a bugfix by
means of a software update is possible and per-
form such an update if necessary.
E1
06
40
Tolerance exceeded: Preheating (VH)
- nom.
z
Perform service routines S005.2
z
Replace the X-ray tube assembly
E1
06
41
Tolerance exceeded: kV - nom.
z
Perform service routines S005.2
z
Replace the X-ray tube assembly
E1
06
42
Tolerance exceeded: mA - nom.
z
Perform service routines S005.2
z
Replace the X-ray tube assembly
E1
06
43
Tolerance exceeded: Preheating (VH)
- act.
z
Perform service routines S005.2
z
Replace the X-ray tube assembly
E1
06
44
Tolerance exceeded: kV - act.
z
Perform service routines S005.2
z
Replace the X-ray tube assembly
E1
06
45
Tolerance exceeded: mA - act.
z
Perform service routines S005.2
z
Replace the X-ray tube assembly
Error code
Description
Actions required
see page
Summary of Contents for ORTHOPHOS XG 3 DS
Page 4: ......
Page 9: ...ORTHOPHOS XG 1General information...
Page 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Page 59: ...ORTHOPHOS XG 2 Messages...
Page 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Page 125: ...ORTHOPHOS XG 3 Troubleshooting...
Page 153: ...ORTHOPHOS XG 4 Adjustment...
Page 269: ...ORTHOPHOS XG 5 Service routines...
Page 433: ...ORTHOPHOS XG 6 Repair...
Page 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Page 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Page 531: ...ORTHOPHOS XG 7 Maintenance...
Page 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...