59 38 399 D3352
5 – 148
D3352.076.01.13.02
07.2008
5.19 Service routine S034
Tabs 5
S034.9: Determining the shadowing limit for C3 and C3F
Quitting the service routine
To return to the Service menu, press the
double arrow key
.
To quit the Service menu and return to the main menu, press the
double arrow key
.
To quit the Service menu and return to the Main menu, press the
up arrow key
above selection field 3.
5.19.6
S034: Test step 9
(XG
Plus
and full version only)
In CEPH programs C3 and C3F for full-version XG
Plus
systems, the shadow-
ing limit set by default can be provided with an individual offset. In this way the
shadowing can be shifted within certain limits.
The offset must be determined by reference to an existing exposure. Starting
from the default setting, an offset can be selected and saved within a range
of +15,000
μ
m to -30,000
μ
m.
For C3 and C3F
Shadowing limit
+15,000
μ
m
-30,000
μ
m
Summary of Contents for ORTHOPHOS XG 3 DS
Page 4: ......
Page 9: ...ORTHOPHOS XG 1General information...
Page 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Page 59: ...ORTHOPHOS XG 2 Messages...
Page 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Page 125: ...ORTHOPHOS XG 3 Troubleshooting...
Page 153: ...ORTHOPHOS XG 4 Adjustment...
Page 269: ...ORTHOPHOS XG 5 Service routines...
Page 433: ...ORTHOPHOS XG 6 Repair...
Page 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Page 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Page 531: ...ORTHOPHOS XG 7 Maintenance...
Page 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...