båÖäáëÜ
59 38 399 D3352
D3352.076.01.13.02
07.2008
5 – 147
Tab 5
5.19 Service routine S034
5.19
S034.8: Determining shadowing limit C1 and C2
Selecting test step 8
1.
Select test step 8 in selection field 2 with the
arrow keys
and confirm your
selection by pressing the
Service key
.
2.
Select the desired offset value using the arrow keys in selection field 1
(step width = 500, offset value = 1000 corresponds to a 1 mm shift of the
shadowing limit).
3.
To save the setting, first press the
Memory key
(R key lights up)
and then press the
R key
.
4.
Take a C1 or C2 exposure to check the effectiveness of the setting
(see operating instructions).
1
2
3
+
-
S034
8
XG
Plus
:
Easypad touchscreen
+
-
(
o
S34
(
n
1
2
3
+
-
5500
XG
Plus
:
Easypad touchscreen
R
(
n
8
R
R
1
2
3
+
-
5500
XG
Plus
:
Easypad touchscreen
R
)
n
)
o
S34
8
Summary of Contents for ORTHOPHOS XG 3 DS
Page 4: ......
Page 9: ...ORTHOPHOS XG 1General information...
Page 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Page 59: ...ORTHOPHOS XG 2 Messages...
Page 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Page 125: ...ORTHOPHOS XG 3 Troubleshooting...
Page 153: ...ORTHOPHOS XG 4 Adjustment...
Page 269: ...ORTHOPHOS XG 5 Service routines...
Page 433: ...ORTHOPHOS XG 6 Repair...
Page 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Page 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Page 531: ...ORTHOPHOS XG 7 Maintenance...
Page 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...