5.3 Service routines with SIDEXIS
5.
3
3.
Select the constancy test in SIDEXIS XG:
Select
EXTRAS
‡
CONSTANCY
TEST
.
The typical Sidexis user interface is started.
Constancy test is already preset.
4.
Start the exposure mode:
Click
XCXP
.
The dialog box for selecting the X-ray device appears on the screen.
5.
Select/confirm the X-ray device:
Select the desired
X
-
RAY
DEVICE
and click
OK
.
The dialog box for selecting the test type appears on the screen.
NOTE
i
Under certain circumstances, the dialog box for selecting the X-ray device may
be skipped by the software and the dialog box for selecting the test type ap-
pears immediately.
6.
Select/confirm the test type:
Click
SERVICE
EXPOSURE
.
The dialog box for selecting the service exposure appears on the screen.
Summary of Contents for ORTHOPHOS XG 3 DS
Page 4: ......
Page 9: ...ORTHOPHOS XG 1General information...
Page 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Page 59: ...ORTHOPHOS XG 2 Messages...
Page 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Page 125: ...ORTHOPHOS XG 3 Troubleshooting...
Page 153: ...ORTHOPHOS XG 4 Adjustment...
Page 269: ...ORTHOPHOS XG 5 Service routines...
Page 433: ...ORTHOPHOS XG 6 Repair...
Page 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Page 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Page 531: ...ORTHOPHOS XG 7 Maintenance...
Page 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...