7.5 Checking the light localizers
7.
5
Checking the position of the light lines
(FH and MS) on the PAN
z
Switch the light localizers on and check the positions of the light lines:
– The
horizontal light line (FH)
must be running horizontally
– The
vertical light line (MS)
must be running vertically through the cen-
ter of the forehead support down to the center of the support piece
NOTE
i
Tip:
When checking and adjusting the light localizers use the following as
reference lines:
For the FH light localizer:
Edge of the tube assembly light localizer
For the MS light localizer: Support piece, inserted bite block or
Forehead support
If the light beam alignment is not correct:
Æ
Adjust the light localizers
(see chapter 6)
Checking the position of the light lines
(FH) on the Ceph
z
Switch the light localizers on and check the positions of the light lines:
Æ
The
horizontal light line (FH)
must be running horizontally between
the temple supports at the height of the ear plugs
NOTE
i
Tip:
When checking or adjusting the light localizer, you may use a PA or AP
view to assess the light beam on the ear plugs.
FH
29
MS
60
90
120
70
80
100 110
FH
Summary of Contents for ORTHOPHOS XG 3 DS
Page 4: ......
Page 9: ...ORTHOPHOS XG 1General information...
Page 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Page 59: ...ORTHOPHOS XG 2 Messages...
Page 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Page 125: ...ORTHOPHOS XG 3 Troubleshooting...
Page 153: ...ORTHOPHOS XG 4 Adjustment...
Page 269: ...ORTHOPHOS XG 5 Service routines...
Page 433: ...ORTHOPHOS XG 6 Repair...
Page 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Page 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Page 531: ...ORTHOPHOS XG 7 Maintenance...
Page 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...