Tab 4
4.3 Adjusting the panoramic X-ray unit
4.
3
Evaluating the X-ray image
7.
Evaluate the X-ray image:
– The superimposed filter must cover one half of the diaphragm (A).
NOTE
i
If this criterion is not yet fulfilled
(B)
, repeat the adjustment procedure starting
with step 5.
If you do not reach your goal via automatic adjustment, repeat the adjustment
procedure with a manually determined adjustment value (see page 4-44).
Saving a value
8.
If the image is identical to the ideal image
(A)
, save the value:
Click
SAVE
VALUES
NOTE
i
In SIDEXIS versions V01.50 and higher, the value for
Fx
in the
PAN
-
FILTER
submenu is set
equal to zero
with a correct adjustment, i.e. if it
is within the permissible tolerance. For versions < V01.50, the value in the text
box may deviate slightly from zero even with a correct adjustment.
z
Go on to the next adjustment step.
Adjustment: ok
A
Summary of Contents for ORTHOPHOS XG 3 DS
Page 4: ......
Page 9: ...ORTHOPHOS XG 1General information...
Page 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Page 59: ...ORTHOPHOS XG 2 Messages...
Page 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Page 125: ...ORTHOPHOS XG 3 Troubleshooting...
Page 153: ...ORTHOPHOS XG 4 Adjustment...
Page 269: ...ORTHOPHOS XG 5 Service routines...
Page 433: ...ORTHOPHOS XG 6 Repair...
Page 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Page 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Page 531: ...ORTHOPHOS XG 7 Maintenance...
Page 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...